MD 30 Operating Instructions, Art. No. 31686, V45/12
7.8
Configuration Menu
In the Configuration Menu the user can customize the device after his favor. The parameters are organized in
different levels. The following information and parameters are stored in there.
• Software version
• Mainboard serial number
• Date and Time
• Display illumination
• Number of programs
• Operating hours MD 30
• Operating hours Motor 1
• Operating hours Motor 2
• Operating hours of irrigation pump
• Error-messages (the last 8)
• Activation of available handpieces
• Speed limitation for each handpiece
• Range of action of AS mode
• Pump behavior
• Motor behavior for 4 different motors
• Behavior of torque at reverse mode
• Reset to factory default
Access to Configuration Menu:
• Press
“Enter”
key for 3 sec. until you hear a long tone.
The right arrow
“
>
”
in the information line on the display indicates that you’re in the Configuration
Menu.
• With
“Select
▲
”
or
“Select
▼
”
choose the desired parameter.
• Push
“Enter”
to go to the sub menu.
• To change the value press
“Select
▲
”
or
“Select
▼
”
• To confirm the changes made, press
“Enter”
for one sec., until you hear a short tone.
• To abort the settings made press
“Enter”
briefly, the setting returns to its previous value.
Exit from Configuration Menu:
• To leave the Configuration Menu press
“Enter”
for 3 sec., until you hear the long tone.
Parameter Level 1
Group/Parameter Rights
Factory
Definition
Software/Version read
VX.XXXX
Shows current software version
Hardware/Serial number MB
read
XXXXXXXXX
Shows serial number of the main board
Date-Time/Date format US
read/write
no
Set to US Date format
Date-Time/Date
read/write
–
Change
current
date
Date-Time/Time read/write
–
Change
current
time
Backlight/brightness (0. .10)
read/write 9
Display
brightness, changeable: 0, … , 10
Programs/Number of programs
read/write
10
Number of activated Programs: 3, … , 10
Operating hours/MD 30
read
0
Shows operating hours of MD 30
Operating hours/Motor 1
read
0
Shows operating hours of motor 1
Operating hours/Motor 2
read
0
Shows operating hours of motor 2
Operating hours/Pump
read
0
Shows operating hours of pump
Error memory/ 1 – 8
read
0
8 Error messages in chronological order.
Be cautious when changing parameters. Unusual behavior of instruments while operating may provoke
false reactions and jeopardize the patient.
Every setting and the new behavior of the instrument has to be verified and tested.