ES2 Electronic Liquid Level Transmitter 

 

 

 

 

 

EX-1845

KING-GAGE ES2

 

Level Transmitter

Two-wire transmitter provides a direct 4-20 mAdc analog electronic output 
signal. The ES2 transmitter is designed specifically for hydrostatic pressure 
measurement for liquid level gauging.

Specifications

Pressure Ranges

0-5, 0-10, 0-15, 0-30, 0-50 psig
(Gage pressure measured relative to ambient atmospheric pressure.)

Accuracy

Less than ± 0.20% FS
Includes linearity, hysteresis, non-repeatability errors.

Pressure Media

Gas or Liquids compatible with 316 stainless steel (diaphragm); 
silicone rubber (O-ring). All other wetted parts 316L.

Electrical

Input (excitation): 14-40 Vdc
Output (mAdc): 4-20 milliamperes 

Span Adjustment Range

Transmitter adjustment turndown is 3:1 for the 0–15, 0–30, 0–50 psig nominal ranges.

Nominal 

Minimum  Range of Adjustment

0–5 psig 

0–3 psig 

0–83 in. thru 0–138 in. water / 0–2.1 m thru 0–3.5 m water

0–10 psig  0–3 psig 

0–83 in. thru 0–277 in. water / 0–2.1 m thru 0–7.0 m water

0–15 psig  0–3 psig 

0–83 in. thru 0–415 in. water / 0–2.1 m thru 0–10.5 m water

0–30 psig  0–10 psig  0–277 in. thru 0–830 in. water / 0–7.0 m thru 0–21.0 m water
0–50 psig  0–15 psig  0–415 in. thru 0–1384 in. water / 0–10.5 m thru 0–35.1 m water

Operating Limit (Maximum Pressure)

Pressure above 300% nominal range (overrange) will result in damage to the transmitter 
(200% may cause a shift in calibration).

Non-Repeatability (Worst Case)

Less than 0.02% FS

Operating Temperature Range

0°F to 175°F/ -17°C to 79°C 

Summary of Contents for KING-GAGE ES2 5501

Page 1: ...sions KING GAGE A NOSHOK Company I 1010 West Bagley Road Berea Ohio 44017 I Ph 855 367 2494 I www king gage com ES2 Electronic Liquid Level Transmitter Effective October 2016 EX 1845 Installation and...

Page 2: ...rchantability are given by this agreement The exclusive remedy for nonconformity of these goods shall be repair and or replacement of the nonconforming goods or parts Seller will not be liable for con...

Page 3: ...ifications 4 Model No Designations 5 Mounting Dimensions 6 7 Typical Application 7 Power Requirement 8 Connectors and Signal Cable 9 10 Test Point Terminals 10 ES2 Calibration 12 13 Calibration Equipm...

Page 4: ...excitation 14 40 Vdc Output mAdc 4 20 milliamperes Span Adjustment Range Transmitter adjustment turndown is 3 1 for the 0 15 0 30 0 50 psig nominal ranges Nominal Minimum Range of Adjustment 0 5 psig...

Page 5: ...r 15 0 15 psig 0 1 03 bar 30 0 30 psig 0 2 06 bar 50 0 50 psig 0 3 44 bar Mounting Application 1 short mount single wall non insulated tank 2 standard mount insulated wall up to 4 5 in 114 mm or adapt...

Page 6: ...nce for installation and removal of the transmitter Clearance dimensions will vary depending upon type of cable or connector used Model 5502 ES2 Slimline Transmitter Standard Mount Model 5501 ES2 Slim...

Page 7: ...19 7 8 9 4 5 6 1 2 3 0 Men u Se lect Ed it En te r Re se t 6 in 152 mm 6 35 in 161 mm 2 Flange 150 lb ANSI Diameter Flange Size ANSI 4 25 108mm 1 pipe flange 150 lb 5 0 127mm 1 5 pipe flange 150 lb 6...

Page 8: ...r and is intended for plug type molded cable assemblies designed for 4 pole signal termination Our ES2 transmitter package also includes a molded M12 cable 32ft 10m to simplify installation and ensure...

Page 9: ...5 cm 25 in 0 5 cm Connector Body Compression Sleeve End Cap Seal Grommet L Transmitter Signal Cable Use twisted pair cable 20 22 AWG for the signal loop between transmitter and receiver 3 The connecto...

Page 10: ...ing the output signal without disturbing the signal loop Do not connect any device or meter whose internal resistance exceeds 20 ohms across these test points Also refer to In Process Zero Span Adjust...

Page 11: ...osts on the transmitter board 1 2 Adjust span potentiometer as necessary to increase output clockwise ordecreaseoutput counterclock wise until the signal displays as 20 00 or 20 01 mA on the test mete...

Page 12: ...m pressure to the sensing diaphragm of the ES2 transmitter requires a pres sure test shell available from King Engineering 2 2 Record the existing setting position of the rotary switch 2 3 Adjust the...

Page 13: ...ion CAUTION Some pressure regulators will allow a slight pressure output even when completely adjusted to the shut position It is preferable to vent the transmitter to atmospherebydisconnectingthepres...

Page 14: ...here A Reserve inches of depth from low point of tank to sensor B Full Tank inches of depth from low point of tank to full C Specific Gravity of Tank Contents MODEL RANGE 5701 1 15 PSIG Sensor Assembl...

Page 15: ...replaced if there are any signs of physical damage such as a cut to the O ring Signal wiring connectors The transmitter provides an M12 threaded outlet on the exterior of the housing that accepts an...

Page 16: ...Page 16 ES2 Electronic Liquid Level Transmitter EX 1845...

Reviews: