ES2 Electronic Liquid Level Transmitter 

 

 

 

 

 

EX-1845

Calibration Equipment Setup

To ensure specified accuracy during field calibration, 
the equipment and setup used must conform to the 
minimum requirements noted in this section.

Milliameter (multimeter)

: 3-1/2 digit minimum, 

0.05% accuracy (internal resistance 20 Ohms or less)

Power Supply: 

24 Volts DC (regulated); permissible 

range is 14-40 Vdc

Pressure Indicator: 

0.05% accuracy (corresponding 

to transmitter nominal range)
0-5 psig 

+ 0.0025 PSI (or 0.07 in. of water)

0-10 psig 

+ 0.0050 PSI (or + 0.14 in. of water)

0-15 psig 

+ 0.0075 PSI (or + 0.21 in. of water)

0-30 psig 

+ 0.0150 PSI (or + 0.42 in. of water)

0-50 psig 

+ 0.0250 PSI (or + 0.69 in. of water)

Pressure Regulator: 

High precision fine thread adjust-

ment for stable air/gas pressure application

CAUTION: Some pressure regulators will allow a slight 
pressure output even when completely adjusted to the 
shut position. It is preferable to vent the transmitter to 
atmosphere by disconnecting the pressure input to ensure 
accurate zero setting of transmitter.

Test Shell Fixture

This pressure fixture is available in the following con-
figurations based on the mounting length of the ES2 
transmitter. Threaded 1/8 NPT connections permit air/
gas pressure to be applied to the face of the transmitter 
when seated within the test shell.

Part No. 

Applicable Transmitter Model

5668-1 

short mount ES2 Model 5501-x-xx

5668-2 

std mount ES2 Model 5502-x-xx

5668-3 

long mount ES2 Model 5503-x-xx 

DC

Power

Supply

Pressure

Indicator

Pressure

Regulator

ES2 Transmitter

installed in Test Fixture 

Shell, p/n 5668-x

Meter Post (-)

Meter Post (+)

Black (-)

Red (+)

Multimeter

2000mA

200mA

20mA

2000

µ

A

DC

mA

1000V DC

750V AC

200V DC

200V AC

20V

2000mA

2000K

200K

20K

2000

V

COM

mA

20.00

Compressed

Air

0

160

Summary of Contents for KING-GAGE ES2 5501

Page 1: ...sions KING GAGE A NOSHOK Company I 1010 West Bagley Road Berea Ohio 44017 I Ph 855 367 2494 I www king gage com ES2 Electronic Liquid Level Transmitter Effective October 2016 EX 1845 Installation and...

Page 2: ...rchantability are given by this agreement The exclusive remedy for nonconformity of these goods shall be repair and or replacement of the nonconforming goods or parts Seller will not be liable for con...

Page 3: ...ifications 4 Model No Designations 5 Mounting Dimensions 6 7 Typical Application 7 Power Requirement 8 Connectors and Signal Cable 9 10 Test Point Terminals 10 ES2 Calibration 12 13 Calibration Equipm...

Page 4: ...excitation 14 40 Vdc Output mAdc 4 20 milliamperes Span Adjustment Range Transmitter adjustment turndown is 3 1 for the 0 15 0 30 0 50 psig nominal ranges Nominal Minimum Range of Adjustment 0 5 psig...

Page 5: ...r 15 0 15 psig 0 1 03 bar 30 0 30 psig 0 2 06 bar 50 0 50 psig 0 3 44 bar Mounting Application 1 short mount single wall non insulated tank 2 standard mount insulated wall up to 4 5 in 114 mm or adapt...

Page 6: ...nce for installation and removal of the transmitter Clearance dimensions will vary depending upon type of cable or connector used Model 5502 ES2 Slimline Transmitter Standard Mount Model 5501 ES2 Slim...

Page 7: ...19 7 8 9 4 5 6 1 2 3 0 Men u Se lect Ed it En te r Re se t 6 in 152 mm 6 35 in 161 mm 2 Flange 150 lb ANSI Diameter Flange Size ANSI 4 25 108mm 1 pipe flange 150 lb 5 0 127mm 1 5 pipe flange 150 lb 6...

Page 8: ...r and is intended for plug type molded cable assemblies designed for 4 pole signal termination Our ES2 transmitter package also includes a molded M12 cable 32ft 10m to simplify installation and ensure...

Page 9: ...5 cm 25 in 0 5 cm Connector Body Compression Sleeve End Cap Seal Grommet L Transmitter Signal Cable Use twisted pair cable 20 22 AWG for the signal loop between transmitter and receiver 3 The connecto...

Page 10: ...ing the output signal without disturbing the signal loop Do not connect any device or meter whose internal resistance exceeds 20 ohms across these test points Also refer to In Process Zero Span Adjust...

Page 11: ...osts on the transmitter board 1 2 Adjust span potentiometer as necessary to increase output clockwise ordecreaseoutput counterclock wise until the signal displays as 20 00 or 20 01 mA on the test mete...

Page 12: ...m pressure to the sensing diaphragm of the ES2 transmitter requires a pres sure test shell available from King Engineering 2 2 Record the existing setting position of the rotary switch 2 3 Adjust the...

Page 13: ...ion CAUTION Some pressure regulators will allow a slight pressure output even when completely adjusted to the shut position It is preferable to vent the transmitter to atmospherebydisconnectingthepres...

Page 14: ...here A Reserve inches of depth from low point of tank to sensor B Full Tank inches of depth from low point of tank to full C Specific Gravity of Tank Contents MODEL RANGE 5701 1 15 PSIG Sensor Assembl...

Page 15: ...replaced if there are any signs of physical damage such as a cut to the O ring Signal wiring connectors The transmitter provides an M12 threaded outlet on the exterior of the housing that accepts an...

Page 16: ...Page 16 ES2 Electronic Liquid Level Transmitter EX 1845...

Reviews: