51
Technical Information
This device is not made with natural rubber latex.
Temperature (Operating)
0 °C to +40 °C (32 °F to +104 °F)
Temperature (Storage/Transportation): -40 °C to +70 °C (-40 °F to +158 °F)
Humidity (Operating)
10% to 95% noncondensing
Humidity (Storage/Transportation): 10% to 95% noncondensing
Altitude (Operating)
Up to 3,000 meters (10,000 feet)
Altitude (Hyperbaric Pressure): Up to 4 atmospheres
Power Requirements
Six 1.5V AA-size alkaline batteries
90 hours - Pulse Oximeter only
20 hours - CO
2
and Pulse Oximeter
24 hours - CO
2
only
Dimensions
8 cm W x 15 cm H x 2.5 cm D
(3 in. W x 6 in. H x 1 in. D)
Weight
310 g (11 oz) (with alkaline batteries)
Classifications per IEC 60601-1 / CAN/CSA-C22.2 No. 601-1 / UL60601-1
Type of Protection: Internally powered (on battery power)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress Protection
IP32