
32
Using the 9847V Pulse Oximeter
Audible alarms re-enabled via
Audible Alarm Disable button or
ending of 2-minute disable
When the audible alarms are re-enabled, and at least one alarm
condition exists, the sound cycle for the highest priority alarm
condition in effect at that time will be started.
Note:
The critically low battery state overrides the audible alarm
disable. If a critically low battery condition occurs and the audible
alarms are either temporarily or permanently disabled, the audible
alarms will become re-enabled. A medium priority alarm will sound
unless there is a high priority alarm condition in effect.
Breath beep
Informational beeps will occur only during the “quiet” part of any
alarm sound in progress (or the beep volume can be set to OFF).
Table 5: Audible Indicator Sound Control Priorities (Continued)
Condition
Sound Control Priorities