Using the 9847V Pulse Oximeter
Using the 9847V Pulse Oximeter
Connecting the Sensors
Pulse Oximeter Sensor
Attach the sensor (with the Nonin logo facing up) to the device (figure 6). Verify the sensor is securely
connected.
Figure 6: Connecting Sensors to the Model 9847V
Carbon Dioxide Sensor and Airway Adapter Tube
Connect the CO
2
sensor (with the Nonin logo facing up) to the side of the device (figure 6). Verify the
sensor is securely attached. Refer to “Carbon Dioxide Sensor” on page 33 for more information.
Turn On/Standby
Press
On/Standby
to turn the device on or off.
Device Startup
When the device is turned on, the device cycles through a startup/initialization sequence before
displaying valid data. During startup, always check for missing indicators or LED display segments. If
any indicator is not functioning, do not use the device. Contact Nonin Technical Service for repair or
replacement.
During its normal startup sequence, the device cycles as follows:
1. Audible breath beep sounds 3 times.
2. No Breath indicator displays for approximately 2 seconds.
3. Low Battery indicator displays for approximately 2 seconds, unless the
batteries are low.
4. Pulse Quality indicator blinks red, then green. The device enters standby if no
pulse oximeter sensor is connected.
Veterinary Pulse
Oximeter Sensor
CO
2
Sensor/Airway
Adapter Tube
No Breath