21
Using the 9847V Pulse Oximeter
Disabling the Audible Alarms During Power On
1. With the unit in Standby, press and hold
Audible Alarm Disable
while pressing and then
releasing
On/Standby
.
2. Release
Audible Alarm Disable
when
888
888
displays on the SpO
2
and pulse rate
displays.
3. After the power on self-test, the Audible Alarm Disabled indicator is steadily lit. The animal
alarms are set at the default limits.
4. The audible alarms can be re-enabled by pressing
Audible Alarm Disable
again.
Setting Parameters
NOTE:
The Advance button sets your selection. The Alarm Limits button moves you to the next
setting.
NOTE:
See page 20 for Table 2 - Alarm Limit, Breath Beep Pitch, Calendar, and Clock Mode
Parameters.
Alarm Mode Setting
1. Upon entering setup mode, “
ALr
” appears in the SpO
2
display indicating alarm mode. The
alarm mode may be set to
dFt
,
rC
L, or
C02
:
•
dFt
(default limits) sets up the system-defined (default) animal alarm limit settings.
•
rC
L (recall stored limits) sets up the last stored animal alarm limit settings of the 9847V.
•
C02
(no breath alarm limits and sensor alarms) temporarily disables the oxygen saturation high
and low settings and the pulse rate high and low settings by setting them to
0FF
. Only the no
breath alarm and the sensor alarms are enabled, along with the breath beep pitch mode
selection. Setup mode skips over the other animal alarm settings (i.e.,
C02
alarm mode first
goes to
nbd
, then to
FPt
, etc.).
2. When the alarm mode setting has been selected, setup mode continues to the no breath
delay time setting.
WARNING: Verify all alarm settings and limits during system startup to ensure that
they are set as intended.
CAUTION:
Review all limits to ensure they are appropriate for the patient.
CAUTION:
Setting alarm limits to extremes can render the alarm system useless.
!
!