©nomics s.a. All rights reserved
MU_Brizzy
+
_NS26_REV 1_EN /09-2019
46/57
Once it has been launched, the calibration process runs automatically. It takes about 20
seconds to complete. The green indicator light on the device flashes rapidly during
calibration.
The probes must be held deep down into the holes of the calibration tool
during the whole calibration process. If a calibration is started when the
probes are not in the proper position, this will result in incorrect mandibular
movement measurement. Recalibrate the device in this case.
During calibration, the calibration tool should stand at a distance of at least 20cm (8") from
any metallic object and from any other electronic device (ex: computer). If you put the
calibration tool down on a table, pay particular attention to the material (possibly metallic)
from which is made the structure supporting the table.