47
CONTROL overview
Control settings
The CONTROL settings are accessible by clicking the gear icon
in the upper
right corner of the Status panel. Clicking the gear icon provides access to the
View
settings menu as shown in
Figure 32
CONTROL settings
Setup
The setup page is shown below in
. Multiple settings for each output
are available and described in the following.
Figure 33
Setup menu page
VIS/NIR RF power
–
or
–
NIR/IR RF power –
Range:
Normal
or
Overdrive
In
Normal
mode, the power level in the main spectral peak increases as RF
power is increased. Power coupling to a fiber deliver system is optimized
when operating in this mode.
Note:
The RF power settings for each crystal are calibrated at the factory together
with the External RF Driver equipped with the system. Unless the crystal or Exter
-
nal RF Driver is changed, this setting should always be set as
Normal
mode.
In
Overdrive
mode, as RF power increases, the main spectral peak only
increases up to a particular level of RF power. After that, increasing RF power
Click the gear icon
to access the menu