5.4.3.
Relay Output
P13: - - - a Relay functions
a
R
ELAY FUNCTIONS
A
LSO SET
:
0
DIFFERENTIAL LEVEL
CONTROL
(Hysteresis control)
Relay is energized if the measured or
calculated value exceeds the value set
in
P14.
Relay is de-energized if the measured
or calculated value drops below the
value set in
P15.
P14
,
P15
There is a need to set
(in level min. 20 mm)
hysteresis between
P14
and
P15
1
ERROR INDICATION
Relay is energized in case of “noECHO”
—
2
Relay is de-energized in case of “noECHO”
—
FACTORY DEFAULT:
P13
= 2
P14: - - - - Relay parameter – Energizing value
P15: - - - -
Relay parameter – De-energizing value
P14
and
P15
parameters must be set when differential level control is required. The parameter values must be set in the unit according to
P01
.
The GET LEVEL function can be used if the parameters affecting the transmitted value are valid. When using differential level control, a minimum
of 20 mm (0.8
″
) hysteresis must be set for the level between the Relay Energizing value and the Relay De-energizing value to avoid frequent
relay state changes.
FACTORY DEFAULT:
P14
= 0,
P15
= 0