20 / 50
sbd31jen21p05
5.3.1.
Basic steps of programming
The unit can be programmed directly with the four push buttons of the SAP–100 module. The followings are an overview of handling the programming module.
Detailed programming instructions are covered in chapters 5.3.2, 5.3.3, 5.3.4, and 5.4.
S
INGLE
-
PRESSING THE BUTTONS
The functions of the keys become active after the button is released.
to save parameter address and go to parameter value; to return from parameter value to parameter address
to move the cursor to the left
to increase the value of the digit under the cursor (press the key continuously to increase the value of the digit faster)
to decrease the value of the digit under the cursor (press the key continuously to decrease the value of the digit faster)
P
RESSING TWO BUTTONS AT THE SAME TIME
Press the two keys simultaneously to access programming steps (0 ... 3 s). The double key combination is marked with ”+” in the following.
E
NTER INTO OR QUIT FROM PROGRAMMING MODES
K
EY COMBINATIONS FOR CHANGING PARAMETER VALUES
GET LEVEL
function can be used only in level and in distance
measurement modes: reading the actual measured value
Notes:
If the blinking cursor cannot be moved to the parameter value from the parameter address after pressing the
button,
the parameter is either read-only, or
modifying the parameter is restricted by a secret code (see
P99
).
If the parameter value keeps blinking the modification cannot be accepted because
the modified value is either out of range, or
the entered code is not valid for this parameter
Summary of Contents for EchoTREK S-300
Page 1: ...sbd31jen21p05 1 50 ...
Page 3: ...sbd31jen21p05 3 50 ...
Page 5: ...sbd31jen21p05 5 50 ...