7. Installation position of sensor
Pipe to the highest point (Gathered bubble
in tube to make error measurement)
Preferred position
(Sensor cannot be installed at highest level of pipeline)
Easily lead to non-full
pipes (wrong measurement way)
Pls put flow meter into low position,
exposure for Injection or discharge
Pls install exhaust valve if
drop pipe longer than 5M.
5
m
Exhaust valve
picture 5 recommend installation
In order to enable reliable flow meter, Pls note the following installation requirements
:
1. As far as possible to avoid ferromagnetic objects and equipment with strong magnetic fields
(big electric machine, big transformer) is protecting of the magnetic field sensor.
2.Flow meter should be enough space around for easy installation and Maintenance.
In longer pipe system, pls install control
valve and Cut-off valve on the flow meter
Downstream side
Pump: Prohibited to install
flowmeter at suction side(Vacuum)