37
FOR HEALTH-CARE PROFESSIONALS
Performance Characteristics
7
Accuracy:
TRUEresult mini accuracy was assessed
against the Yellow Springs Instrument (YSI) Model
2300. Studies were conducted at 3 clinical sites by
Health-care Professionals.
Fingertip Capillary Blood
*40EBUBĕOHFSWT:4*
4
/
Slope 0.98 y-intercept -0.03 mmol/L
r 2 0.98
Range 1.3 - 27.7 mmol/L
7FOPVT#MPPE/
Slope 0.97 y-intercept 0.52 mmol/L
r 2 0.98
Range 2.0 - 29.2 mmol/L