18
11.
Compare result to Control
Solution Range printed on
Strip vial label for Control
Solution Level you are using.
If result is in range,
System can be used
for testing blood. If
result does not fall
within range, repeat
test using a new Strip.
Note:
Control Test result shows the Control Symbol in
the Display.
If Control Test result is outside range, test again.
If result is still outside range, System should not
be used for testing blood. Call for assistance.
After result is shown, hold Meter with
Test Strip pointing down. Press Strip
Release Button to release and discard
Strip in appropriate container. Meter
turns off.
Note:
Removing Strip before result
displays cancels the test. An error
message appears and the result is not stored in
Memory. Retest with a new Strip and do not
remove before result is displayed.
Test Strip
Vial Label
Control Solution
Bottle Label