How to Test Control Solution
1.
Allow Control Solution, vial of Strips
and Meter to adjust to 20°C-25°C
(room temperature). Write date first
opened on Control Solution bottle
label and Strip vial label.
Note:
Running a Control Test at temperatures
outside the range listed above may cause
Control Solution to read as a blood test.
Check dates on Control Solution label and Strip
vial label. Do not use Control Solution or strips
if Use By Date have passed (Control Solution
- 3 months after opening or date next to on
label, Strips - 4 months after opening or date
next to on label). Discard expired products
and use new products.
3.
Swirl or invert bottle gently to mix Control
Solution.
DO NOT SHAKE
!
4.
Remove one Strip from vial. Close Strip vial
immediately.
Note:
Use Strip quickly after removal from vial.
5.
Insert Strip into Test Port. Meter turns on.
Note:
If Strip has been out of the vial
too long before testing, an error
message appears upon insertion
of the Strip into the Meter.
Release and discard old Strip.
Use new Strip.