Cautions
This electrophoresis device must be operated together with an external DC power
supply designed specifically for electrophoresis applications under a high voltage. The
output of this power supply must be isolated from external ground to insure that the
DC voltage output floats with respect to ground.
If this device is used or modified in a manner not specified in this manual, then
protection afforded by the device will be impaired. Alteration of this device will
create a potential safety hazard.
Nippon Genetics Europe GmbH is not responsible for any injury or damage caused by
use of this device for purposes other than those for which it is intended or by
modifications of the device.
The electric operating parameters for this device are as follows:
Voltage Limit: 0 - 600 Vdc
Current Limit: 0 - 100 mA
Ambient Temperature Limit: 0
–
50°C
Protective gloves and safety glasses must be worn when operating in a laboratory
environment. This device is intended for in vitro use only and must be used with its
Lid whenever it is operated.
Precast Gel Cassettes that are separately available (see ordering information) are
breakable if pressed improperly when attached or detached. Gel Cassettes should be
set again into the device when difficult to be assembled properly.
Limited Warranty
This device is warranted against manufacturing defects for a period of one year from
the original purchase date. Nippon Genetics Europe will repair/exchange the
defective parts under this warranty subject to the terms and conditions of Nippon
Genetics Europe, provided that the device is operated according to the manual and
labels.