Nippon FastGene FG-02 Instruction Manual Download Page 3

 

 

 
 

Features 

o

 

Unique wedge system to firmly and easily fix the Gel Cassette(s) 

o

 

Buffer filled in the Chambers to cool all the surface of the Gel Cassettes(s) 

o

 

Smaller quantity of buffer required 

o

 

Easier sample loading visible through the transparent body 

 

Components 

o

 

1x Lid with power cables (1) 

o

 

1x Anode Buffer Chamber (2) 

o

 

1x Cathode Buffer Chamber Frame with electrodes (3) and Gasket (4) 

o

 

2x blue Wedges (5) 

o

 

2x Dummy Gel Cassettes (6) 

o

 

1x Instruction Manual 

 

 

Specifications 

o

 

Lid Acrylic 

 

 

 

Acrylic 

o

 

Anode Buffer Chamber Frame  

Acrylic 

o

 

Cathode Buffer Chamber Frame 

Acrylic 

o

 

Gasket   

 

 

 

Silicone 

o

 

Wedge   

 

 

 

Acrylic 

o

 

Overall size (L x W x H) 

 

8x18x18 cm 

o

 

Weight (before buffer filled)   

0.5 kg 

o

 

Precast Gel: Compatibility   

IDMax

 Gels 

10x8 cm

 (see ordering information) 

o

 

Electric Limits   

 

 

600 Vdc and 100 mA 

o

 

Buffer Camber Capacity  

 

500 ml 

 
This device is not compatible with acetone or ethanol. Use of organic solvents voids 
the warranties. 
 
Additional equipment and materials required (not included): 
Power Supply, Electrophoresis Buffer, Sample Loading Solution. 
 

Summary of Contents for FastGene FG-02

Page 1: ...For Research Use Only For safety this manual must be read carefully and the device be handled properly FastGene Protein Electrophoresis Chamber Instruction Manual...

Page 2: ...rent Limit 0 100 mA Ambient Temperature Limit 0 50 C Protective gloves and safety glasses must be worn when operating in a laboratory environment This device is intended for in vitro use only and must...

Page 3: ...Gel Cassettes 6 o 1x Instruction Manual Specifications o Lid Acrylic Acrylic o Anode Buffer Chamber Frame Acrylic o Cathode Buffer Chamber Frame Acrylic o Gasket Silicone o Wedge Acrylic o Overall si...

Page 4: ...athode Buffer Chamber Frame Otherwise place the Gaskets on the grooves of the Inner Frame and press the Gaskets so that they are fixed firmly into the grooves 3 Pour the buffer approximately 200 ml in...

Page 5: ...ng of the level of the inside buffer If leakage is observed disassemble the inner Chamber and repeat the above steps Note Avoid a short circuit do not overfill the Inner Chamber 7 Fill the Outer Chamb...

Page 6: ...ches approximately 5 mm around the center of Gasket from the bottom edge of the Gel Plate 14 Disconnect the power cables of the Lid for the power supply Holding the device firmly side the Lid as indic...

Page 7: ...ne Protein Electrophoresis Chamber FG 02 IDSolTM Running Buffer 10x Running Buffer Tris Glycine SDS 500 ml ID1501 10x Running Buffer Tris Glycine 500 ml ID1511 10x Running Buffer Tris Tricine SDS 500...