KFISH/01/19.11.2018 

                                                   

 

13.

 

EMC (Electromagnetic Compatibility) Decalaration 

This equipment generates, uses and can radiate radio frequency (RF) energy. This equipment 

may cause electromagnetic interference if it is not installed and used in accordance with the 

manual. 

This device has been tested for Medical Products in accordance with EN-60601-1-2 and has 

been determined to comply with acceptable limits. These limits indicate that the device 

provides an acceptable level of protection against electromagnetic interference (EMC) if the 

device is used as specified in the manual. 

This device may be affected by portable and mobile RF communication devices. This 

equipment must not be stored together with other equipment. 

See the following tables for more information about this device and EMC. 

 

Guidance and Manufactures Declaration-Electromagnetic 

 

These devices are intended for use in the electromagnetic environment specified below. The customer or the user of 
the device should assure that the device is used in such an environment. 

Emissions Test  Compliance 

Electromagnetic environment

guidance 

 

 

 

RF emissions 
CISPR 11  

Group 1 

The device uses RF energy only for its internal function. Therefore, its RF 
emissions are very low and 

 

 

are not likely to cause any interference in nearby electronic equipment. 

 

 

 

 

 

 

RF emissions 
CISPR 11 

Class B 

The device is suitable for use in all establishments, including domestic 
establishments and those directly connected to the public low-voltage network 
that supplies buildings used for domestic purposes. 

 

 

   

 

 

 

Harmonic 
Emissions 

Class A 

 

 

 

 

   IEC 61000-3-2 

 

 

 

 

 

Voltage 
Fluctuations/Flick
er Emissions IEC 
61000-3-3 

Complies 

 

 

 

 

 

 

 

 

 

 

 

 

 
 
 
 
 

17 

                                                                                                                           KFISH/01/19.11.2018 

                                                   

 

 

 

Guidance and Manufactures Declaration-Electromagnetic 

 

These devices are intended for use in the electromagnetic environment specified below. The customer or the user of 
the device should assure that the device is used in such an environment. 

Immunity Test 

IEC60601 Test 
Level 

 Compliance      
Level 

Electromagnetic environment

guidance 

 

 

 

 

Electrostatic 
discharge 
(ESD) IEC 61000-
4-2 

± 6 kV contact 

± 6 kV contact 

Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should be 
at least 30%. 

 

 

± 8 kV air 

± 8 kV air 

 

 

 

 

 

Electrical fast 
transient/ 
burst IEC 61000-4-

±2 kV for power 
supply lines 
 
±1 kV for 
input/output lines 

± 6 kV for power 
supply lines 

 
 
± 1 kV for 
input/output lines 

Mains power quality should be that of a typical commercial or 
hospital environment. 

 

 

Surge 

±1 kV differential 
Mode 
 
±2 kV common 
mode 

±1 kV differential 
mode 
±2 kV common 
mode 

Mains power quality should be that of a typical commercial or 
hospital environment. 

 

IEC 61000-4-5 

 

 

 

 

Voltage dips, short 
interruptions and 
voltage variations 
on power supply 
input lines 
IEC 61000-4-11 

  <5 % UT 

(>95 % dip in UT) 

for 0,5 cycle 

40 % UT 

(60 % dip in UT) 

for 5 cycles 

70 % UT 

(30 % dip in UT) 

for 25 cycles 

<5 % UT 

(>95 % dip in UT) 
for 5 sec

 

<%5 

U

(

U

T

’de >%95

 

düşüş) 0,5 devir

 

için 

%40 

U

(

U

T

’de %60 düşüş)

 

5 devir için 

%70 

U

(

U

T

’de %30 düşüş)

 

25 devir için 

<%5 

U

(

U

T

’de >%95

 

düşüş) 5 saniye

 

için

 

  Mains power quality should be that of a typical commercial or         
hospital 

environment. 
If the user of the device requires continued operation during 
power 
mains interruptions, it is recommended that the device be 
powered from 
an uninterruptible power source 

 

 

 

 

 

 

 

 

 

 

Power frequency 
(50/60Hz) 
magnetic field 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical commercial or 
hospital environment. 

 

 

 

 

 

 

 

 
 

 

 

 

 

Note:

 Ut is the AC mains voltage prior to application of the test level. 

 
 

 

 

18 

                                                                                                                           KFISH/01/19.11.2018 

                                                   

 

 

Guidance and Manufactures Declaration-Electromagnetic 

 

These devices are intended for use in the electromagnetic environment specified below. The customer or the user 
of the device should assure that the device is used in such an environment. 

Immunity Test 

IEC60601 Test  
Level 

Compliance 
Level 

Electromagnetic environment

guidance 

 

 

      

 

 

Conducted RF 
IEC 61000-4-6 

3 V rms 

3 V 

Portable and mobile RF communications equipment should 
be used no 
closer to any part of the device, including cables, than the 
recommended 
separation distance calculated from the equation applicable 
to the 
frequency of the transmitter 
 
Recommended separation distance 

150 kHz ila 80 MHz 

 

Radiated RF 
IEC 61000-4-3 

 

 

 

 

10 V rms 

10 V 

 

 

150 kHz ila 80 MHz 

 

 

 

 

 

  

 

 

10 V/m 

80 MHz ila 2,5 GHz 

 

10 V/m

 

 

 

 

 

   

 

 

 

 

Where (P) is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter manufacturer 
and d is the recommended separation 
distance in meters (m). Field strengths from 
fixed RF transmitters, as determined by an 
electromagnetic site survey, a should be less 
than the compliance level in each frequency 
range. b Interference may occur in the 
vicinity of equipment marked with the 
following symbol:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

      

 

Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. 

 

Not 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people. 

 

a.  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, 

amateur radio,AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the 
electromagnetic environmentdue to fixed RF transmitters, an electromagnetic site survey should be considered. If the 
measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the 
device should be observed to verify normal operation. If abnormalperformance is observed, additional measures may be 
necessary, such as reorienting or relocating the device. 

b.  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m 

 

 

 
 
 
 
 
 
 

19 

                                                                                                                           

KFISH/01/19.11.2018 

                                                   

 

 

  Recommended separation distance between portable and mobile RF communication equipment and this device 

 

 

宗宫宨家宨季宧宨容宬宦宨家季室宵宨季宬宱宷宨宱宧宨宧季宩宲宵季宸家宨季宬宱季室宱季宨宱容宬宵宲宱宰宨宱宷季宬宱季宺宫宬宦宫季宵室宧宬室宷宨宧季宕安季宧宬家宷宸宵宥室宱宦宨家季室宵宨季宦宲宱宷宵宲宯宯宨宧孱季宗宫宨季宦宸家宷宲宰宨宵季宲宵季宷宫宨季
宸家宨宵季宲宩季宷宫宨季宧宨容宬宦宨季宦室宱季宫宨宯害季害宵宨容宨宱宷季宨宯宨宦宷宵宲宰室宪宱宨宷宬宦季宬宱宷宨宵宩宨宵宨宱宦宨季宥宼季宰室宬宱宷室宬宱宬宱宪季室季宰宬宱宬宰宸宰季宧宬家宷室宱宦宨季宥宨宷宺宨宨宱季害宲宵宷室宥宯宨季室宱宧季
宰宲宥宬宯宨季宕安季宦宲宰宰宸宱宬宦室宷宬宲宱家季宨宴宸宬害宰宨宱宷季孫宷宵室宱家宰宬宷宷宨宵家孬季室宱宧季宷宫宨季宧宨容宬宦宨季室家季宵宨宦宲宰宰宨宱宧宨宧季宥宨宯宲宺孯季室宦宦宲宵宧宬宱宪季宷宲季宷宫宨季宰室宻宬宰宸宰季宲宸宷害宸宷季
害宲宺宨宵季宲宩季宷宫宨季宦宲宰宰宸宱宬宦室宷宬宲宱家季宨宴宸宬害宰宨宱宷孱

 

Rated maximum 
output power 
of transmitter (W) 

Separation distance according to frequency of 
transmitter (m) 

 

   

 

   

 

 

 

 

 

 

 

 

 

 

 

 

150 kHz ila 

150 kHz ila 

80 MHz ila 

 

 

80 MHz ila 

   

80 MHz, Outside ISM 
Bands  

80 MHz,In ISM 
Bands 

800 MHz 

 

 

800 MHz 

   

 

 

 

 

 

 

 

 

 

 

   

 

   

 

 

   

 

   

 

   

 

   

 

 

   

 

   

 

 

 

   

0,01 

0,12 

   

0,12 

   

0,12 

 

0,23 

   

 

 

 

 

 

 

 

 

 

 

 

 

0,1 

0,37 

   

0,38 

   

0,38 

 

0,73 

   

 

 

 

 

 

 

 

 

 

 

 

 

1,16 

   

1,20 

   

1,20 

 

2,30 

   

 

 

 

 

 

 

 

 

 

 

 

 

10 

3,67 

   

3,79 

   

3,79 

 

7,27 

   

 

 

 

 

 

 

 

 

 

 

 

 

100 

11,60 

   

12,00 

   

12,00 

 

23,00 

   

 

 

 

 

 

 

 

 

 

 

 

 

 

 

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres           
(m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output 

    power rating of the transmitter in watts (W) according to the transmitter manufacturer. 

 

Note 1

. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.. 

 

Note 2

. ISM (industrial, scientific, and medical) bands from 150 kHz to 80 MHz are 6,765 MHz to 6,795 MHz; 

13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz and 40,66 MHz to 40,70 MHz.

 

 

Note 3

. These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 

 reflection from structures, objects and people.  

 

 

Summary of Contents for HNK-NBL-FISH

Page 1: ...i in uygundur Ancak bilinci a k olmayan veya kendili inden soluyamayan hastalarda kullan m i in uygun de ildir Bebekler k k ocuklar ve zel bak m gerektiren ki ilerin cihaz yeti kin ki i kontrol alt nd...

Page 2: ...lar kullan n z Orijinal aksesuarlar d nda ve kullan c n n ki isel sebeplerinden kaynaklanan hasarlar i in garanti hizmeti sa lanmaz Cihaz kullanmadan nce t m aksesuarlar n al r vaziyette oldu undan em...

Page 3: ...la kab ila kab kapa ve a zl k par alar n kaynar su i inde 10 dakika bekletin 3 Yeti kin maskesi ve ocuk maskesi ise sirke saf su kar m i erisinde tamamen bat r lacak ekilde 30 dakika bekletilir 4 Daha...

Page 4: ...kullan l p kullan lmad n kontrol ediniz F Hava hortumun katlan p katlanmad n kontrol ediniz F Cihaz hava filtresi t kanm olabilir kontrol ediniz Cihaz a r derecede s n yor F Cihaz havaland rma kanall...

Page 5: ...z aras ndaki ISM frekans bantlar n n ve 80 MHz ila 2 5 GHz aras ndaki frekans aral n n uygunluk seviyeleri mobil ta nabilir ileti im ekipmanlar n n yanl l kla hastan n bulundu u alana getirildi inde e...

Page 6: ...3 EMC Electromagnetic Compatibility Decalaration 16 14 Symbols and Descriptions 20 15 Device Outside Box s Symbols Descriptions 21 16 European Conformity Mark 21 17 Manufacturer and Technical Service...

Page 7: ...ieces which can go down easily This device shouldn t use in the enviroment where inflammable gases oxygen and aerosol sprays used in This device is used for breathing passage treatment under the contr...

Page 8: ...r which is not boiled and clean the pieces Less than 50 C Don t wash the device s main unit air tube and air filter 1 Take out the device s mouthpiece mask medication cup and air tube 2 Empty the resi...

Page 9: ...4 Test 5 Test 14 KFISH 01 19 11 2018 10 Keeping and Transporting The device can be kept with its package in well aired room humid is not more than 10 95 and the temperature between 25 70 C Under this...

Page 10: ...uld be at levels characteristic of a typical location in a typical commercial or hospital environment Note Ut is the AC mains voltage prior to application of the test level 18 KFISH 01 19 11 2018 Guid...

Page 11: ...ox s Symbols Descriptions Warning Caution Note Keep Away From Sunlight Keep Away From Getting Wet Handle With Care Rohs Declaration Barcode Number 16 European Conformity Mark Notified Body KIWA Belgel...

Reviews: