IV Shutter Control Unit Communications Commands
1 Communication-controlled operation
- 32 -
1-7 Commands
(1) c command: Operation, control, and data setting command
c command
o: Normal response
n: Error response
Description of operation
cTSO(1)
oTSO
nTSO(6)
Opens a specified shutter and closes any other shutters. If an
open time period has been specified, the shutter will close
when the specified period elapses. Note that no shutters will
open if the eyepiece shutter is open.
(1): [1-3] Shutter number
(6): Error code
cTSC
oTSC
nTSC(6)
Closes all shutters.
cTTS(2):(3)
oTTS
nTTS(6)
Specifies a shutter opening time.
(2): [1-3] Shutter number
(3): [0000-1000] Time setting
* [0000] indicates no opening time. Unit: 0.1 sec (100.0 sec
max.)
(6): Error code
cTSW(4)
oTSW
nTSW(6)
Enables/disables the laser open/close switch on the control
panel.
(4): [1] Enable, [0] Disable
(6): Error code
(2) r command: Data request command
r command
a: Normal response
n: Error response
Description of operation
rTSR
aTSR(1)
nTSR(6)
Calling of whether any shutter is open.
(1): [0] All shutters closed, [1-3] Number of open shutter
(6): Error code
rTTS(1)
aTTS(2)
nTTS(6)
Calling of shutter opening time setting.
(1): [1-3] Shutter number
(2): [0000-1000] Setting
(6): Error code
rTES
aTES(3)
nTES(6)
Calling of whether the eyepiece shutter is open or closed.
(3): [1] Open, [0] Closed
(6): Error code
rTSW
aTSW(4)
nTSW(6)
Calling of whether the laser open/close switch on the control
panel is enabled or disabled.
(4): [1] Enabled, [0] Disabled
(6): Error code
rVER
aVER(5)
nVER(6)
Calling of program version.
(5): V*.** (five characters with "*" being numbers. Ex.: V1.00)
(6): Error code
Summary of Contents for TIRF