Troubleshooting

Specifications

Cause

Problem

Solution

To measure the coating thickness for a 
steel plate with a zinc coating, and a film 
coating on top, Using Automatic Mode, first 
measure the thickness of both the zinc 
layer and the top coating. Next, switch to 
NFE measurement mode and measure the 
thickness of the film. The thickness of the 
zinc layer is the difference between the two 
measurements. 

<

Example

>

Instrument is able to use both magnetic induction for measuring 
thickness of coating on iron and steel, and eddy current method 
for measuring thickness of coating on non-ferrous material. 
Instrument can automatically select method of measurement.

Just NFE-

MEASUREMENT

AUTOMATIC

MEASUREMENT

BACK

NFE Measurement Mode :

Uses eddy current measurement method.

Auto Mode :

Automatically select magnetic induction, or eddy 
current measurement method.

Press and Hold to return to Main Menu.

LANGUAGE

BATTERY

TURNOFF-TIME

INSTR.RESET

BACK

Select Language : 

Toggle display from English to German.

Battery Status :

Display of current battery voltage. Instrument 
turns off if voltage drops below 1.0V, so please 
replace. battery.

Auto-Off :

Switchable between 3 options: 1 min. (default) / 
5 min. / 30 min. 
Battery consumption will be high if set to 30 min., 
so please only use this setting as necessary.

Setting Initialization :

Use when calibration can not be properly 
performed, or if malfunction occurs.

Press and Hold to return to Main Menu.

2 Setup Menu

Coating

Zinc (non-ferrous material)

Steel Plate (ferrous)

Increment Mode :

Short press to increase the value. Hold 
for 1 sec. to switch to decrement mode.

Decrement Mode :

Short press to decrease the value. Hold 
for 1 sec. to switch to increment mode.

FOIL

ADJUST

Calibration Value Adjust Menu Item :

This mode is used to set the reading to the 
thickness of the Calibration-foil. Typically, the 
thick Test Film is used for calibration 

(Calibration-foil), but if only measuring thinner 
films, the thin Test Film can be used (Control-foil).

Foil Value >

295

μ

m

Foil Value <

295

μ

m

FOIL

CALIBRATION

BACK

If pressed 10 times in one direction, dir. 
control will roll over.
When reading matches the thickness 
of the Calibration-foil, Press and hold 
to set the value.

Press and Hold to return to Main Menu.

Press and hold to switch to 2-Point 
Calibration Menu Item.

SETUP

MODE

Long Press

Long Press

Long Press

Long Press

2 Measurement Mode Menu (FTM-52FN only)

Battery is low.

Please replace battery.

Please correct battery orientation. 

Please initialize meter.

Using FTM-50FE 

on non-ferrous 

substrate material.

Battery is inserted 
with wrong polarity.

Probe is not making 
good contact.

Coating is too thick.

Fault has occurred.

FTM-50FE only measures coating on iron 
based material. For non-ferrous substrate, 
please use the FTM-52FN.

Please press the probe face firmly against 
the surface to be measured.

If coating thickness varies it may be 
above measurement range and therefore 
impossible to measure. 

Substrate material 
has low magnetic 
reactance.

Measurements can not be made on 
substrate with weak ferro-magnetisim, 
such as stainless steel.

Calibration is not 
correct.

Please re-calibrate, and make sure probe 
face is firmly pressed against 
measurement surface.

Surface is rough.

Take several measurements at different 
positions and average the result.

Surface is uneven.

Take several measurements at different 
positions and average the result.

Measurement at 
edge of surface.

Magnetic flux is uneven near the edge of 
a surface; Please take measurement 
away from the edge by at least 15mm. 

Temperature shift 
between calibration 
and measurement.

Calibrate the instrument at the same 
temperature as the measurements.

Surface has 
residual magnetism.

Coating deforms 
when being 
measured.

Place Calibration-foil on top of surface, 
measure, and then subtract the thickness 
of Calibration-foil for correct value.

Probe or Calibration 
film was dirty during 
calibration.

Please remove any contamination and 
recalibrate.

If surface has magnetic field, it may affect 
measurements. Please demagnetize.

Measurements 
have large 
variation.

There is no 
response 
when probe 
is applied to 
surface.

Power does 
not turn On.

Model No.
Measurement Method
  

 

Measurement Range
 

 

Resolution 

Accuracy  

Power Supply
Continuous on time
Probe Material
Size
Weight
Guaranteed Operational Temp.
Storage Temp.
Accessories

After use maintenance and storage

Battery

If instrument will not be used for a long period, please remove 
the battery.

Battery Replacement

 

Replace the battery if 

Change BAT.!

 message is displayed. 

Instrument shuts down if battery voltage drops below 1.0V. 

Storage

After use, wipe instrument with a dry cloth and place in the 
storage case. Store out of direct sunlight in location with 
temperature between 0 and 45

°

C. 

FTM-52FN / FTM-50FE
Eddy Current
Electromagnetic Induction 
Eddy Current 

: 0-2000

μ

m

Electromagnetic Induction : 0-5000

μ

100

μ

: 0.1

μ

m

1000

μ

: 1

μ

1.00mm  : 0.01mm

300

μ

:

±

2

μ

m

300 - 1000

μ

m  

:

±

1%

1000 - 2000

μ

m   :

±

3%

2000

μ

:

±

5%

(1x) AA size Battery (Sold Separately)
30 hr. 
Ruby
approx. 

Ø

28 

×

 98mm

approx. 72g
0~45

°

C

0~45

°

C

Reference Base (Iron, Non-Ferrous

)

Test Film (2x)  

Storage Case     

Instruction Manual

 Included only with FTM-52FN