I335-K 1707
Troubleshooting
Specifications
Cause
Problem
Solution
To measure the coating thickness for a
steel plate with a zinc coating, and a film
coating on top, Using Automatic Mode, first
measure the thickness of both the zinc
layer and the top coating. Next, switch to
NFE measurement mode and measure the
thickness of the film. The thickness of the
zinc layer is the difference between the two
measurements.
<
Example
>
Instrument is able to use both magnetic induction for measuring
thickness of coating on iron and steel, and eddy current method
for measuring thickness of coating on non-ferrous material.
Instrument can automatically select method of measurement.
Just NFE-
MEASUREMENT
AUTOMATIC
MEASUREMENT
BACK
NFE Measurement Mode :
Uses eddy current measurement method.
Auto Mode :
Automatically select magnetic induction, or eddy
current measurement method.
Press and Hold to return to Main Menu.
LANGUAGE
BATTERY
TURNOFF-TIME
INSTR.RESET
BACK
Select Language :
Toggle display from English to German.
Battery Status :
Display of current battery voltage. Instrument
turns off if voltage drops below 1.0V, so please
replace. battery.
Auto-Off :
Switchable between 3 options: 1 min. (default) /
5 min. / 30 min.
Battery consumption will be high if set to 30 min.,
so please only use this setting as necessary.
Setting Initialization :
Use when calibration can not be properly
performed, or if malfunction occurs.
Press and Hold to return to Main Menu.
□
2 Setup Menu
Coating
Zinc (non-ferrous material)
Steel Plate (ferrous)
Increment Mode :
Short press to increase the value. Hold
for 1 sec. to switch to decrement mode.
Decrement Mode :
Short press to decrease the value. Hold
for 1 sec. to switch to increment mode.
FOIL
ADJUST
Calibration Value Adjust Menu Item :
This mode is used to set the reading to the
thickness of the Calibration-foil. Typically, the
thick Test Film is used for calibration
(Calibration-foil), but if only measuring thinner
films, the thin Test Film can be used (Control-foil).
Foil Value >
295
μ
m
Foil Value <
295
μ
m
FOIL
CALIBRATION
BACK
※
If pressed 10 times in one direction, dir.
control will roll over.
When reading matches the thickness
of the Calibration-foil, Press and hold
to set the value.
Press and Hold to return to Main Menu.
Press and hold to switch to 2-Point
Calibration Menu Item.
SETUP
MODE
Long Press
Long Press
Long Press
Long Press
□
2 Measurement Mode Menu (FTM-52FN only)
Battery is low.
Please replace battery.
Please correct battery orientation.
Please initialize meter.
Using FTM-50FE
on non-ferrous
substrate material.
Battery is inserted
with wrong polarity.
Probe is not making
good contact.
Coating is too thick.
Fault has occurred.
FTM-50FE only measures coating on iron
based material. For non-ferrous substrate,
please use the FTM-52FN.
Please press the probe face firmly against
the surface to be measured.
If coating thickness varies it may be
above measurement range and therefore
impossible to measure.
Substrate material
has low magnetic
reactance.
Measurements can not be made on
substrate with weak ferro-magnetisim,
such as stainless steel.
Calibration is not
correct.
Please re-calibrate, and make sure probe
face is firmly pressed against
measurement surface.
Surface is rough.
Take several measurements at different
positions and average the result.
Surface is uneven.
Take several measurements at different
positions and average the result.
Measurement at
edge of surface.
Magnetic flux is uneven near the edge of
a surface; Please take measurement
away from the edge by at least 15mm.
Temperature shift
between calibration
and measurement.
Calibrate the instrument at the same
temperature as the measurements.
Surface has
residual magnetism.
Coating deforms
when being
measured.
Place Calibration-foil on top of surface,
measure, and then subtract the thickness
of Calibration-foil for correct value.
Probe or Calibration
film was dirty during
calibration.
Please remove any contamination and
recalibrate.
If surface has magnetic field, it may affect
measurements. Please demagnetize.
Measurements
have large
variation.
There is no
response
when probe
is applied to
surface.
Power does
not turn On.
Model No.
Measurement Method
Measurement Range
Resolution
Accuracy
Power Supply
Continuous on time
Probe Material
Size
Weight
Guaranteed Operational Temp.
Storage Temp.
Accessories
After use maintenance and storage
●
Battery
If instrument will not be used for a long period, please remove
the battery.
●
Battery Replacement
Replace the battery if
“
Change BAT.!
”
message is displayed.
Instrument shuts down if battery voltage drops below 1.0V.
●
Storage
After use, wipe instrument with a dry cloth and place in the
storage case. Store out of direct sunlight in location with
temperature between 0 and 45
°
C.
FTM-52FN / FTM-50FE
Eddy Current
Electromagnetic Induction
Eddy Current
: 0-2000
μ
m
Electromagnetic Induction : 0-5000
μ
m
100
μ
m
: 0.1
μ
m
1000
μ
m
: 1
μ
m
≧
1.00mm : 0.01mm
<
300
μ
m
:
±
2
μ
m
300 - 1000
μ
m
:
±
1%
1000 - 2000
μ
m :
±
3%
≧
2000
μ
m
:
±
5%
(1x) AA size Battery (Sold Separately)
30 hr.
Ruby
approx.
Ø
28
×
98mm
approx. 72g
0~45
°
C
0~45
°
C
・
Reference Base (Iron, Non-Ferrous
※
)
・
Test Film (2x)
・
Storage Case
・
Instruction Manual
※
Included only with FTM-52FN