background image

Before using for measurements, please perform 0-Point 
Calibration, and 2-Point Calibration.

Safety Precautions 

Preparing to Use 

Please Observe

To prevent injury to yourself and others, and to prevent 
damage to property, always follow the procedures below 
marked with the following symbols. 

To prevent accident or injury, please observe the 

information identified with the following notice.

Read this manual and follow the instructions.

Use of product other than as described in the manual 
may cause accident.

Do not get wet.

It may cause failure.

Handle with Care.

Do not drop or shock product. Do not place under 
heavy object or subject product to strong forces. 
Improper use may case inaccuracy or damage.

Handle probe head with care; do not use to strike 
surface as it will cause failure.

Only use to measure applicable materials.

  

Do not disassemble or modify.

It may cause damage or inaccuracy.

Keep away form high voltage and strong 
magnetic fields.

May cause damage or inaccuracy.

Do now wipe with solvent, or solvent based 
products, such as insecticide, etc.

It may cause failure.

Remove battery when not in use for long 
periods.

Keep away from direct sunlight and high 
temperatures such as in a car, or near a stove or 
heat source.

It may cause failure.

After use, wipe off any dirt or dust, and store in 
dust free location.

Installing battery

Battery not included. 
Requires one AA size alkaline battery.

Do not move probe during measurement. Only 
measure single point at a time.

Continuous measurements will cause the probe to 
heat up, so please remove from surface and allow 
probe to cool in air for 

about 2 seconds

 between 

each measurement.

Always remove any dirt from probe surface before 
measuring,

Alkaline AA Battery

Battery Cover

Open

Close

How To Use

Basic Operation

Turn on Power

Press and Hold the ON/OFF

MENU button until 

READY

 

appears on the LCD Display.

 ※

 Auto-Off feature will turn off power after being idle for set 

period. (Initial setting is 1 minute.)

Measure

Press Probe face against surface, when indicator sound is 
heard, measurement is complete. Thickness reading is 
displayed on the LCD display.

 

Iron and Steel (FE)  : 1 beep

 

Non-ferrous (NFE)   : 2 beeps

Turn off Power

Press and Hold the ON/OFF

MENU button until LCD display 

turns off.

Long Press

Power ON/OFF

MENU Button

For an accurate measurement, 0-Point and 2-Point Calibration 
must first be performed using a Reference Base.
Before taking measurement, 

always

 calibrate using uncoated 

substrate of same material as substrate used in the 
measurement. 
When measuring uneven shapes and pipes, calibrate using 
uncoated substrate of same material and shape as the surface 
to be tested.

Calibration Note

Rotate Battery Cover to open, Insert AA size alkaline battery in 
direction shown. Please check orientation before replacing 
cover.

LCD Display

Press face against surface

Probe

Calibration Set Value must match the thickness of 
the Calibration-foil. If they do not match, change the 
setting using the  

“■

Changing The Preset 

Calibration Value

 procedure below. 

Calibrate using calibration film of slightly larger 
thickness than the coating you will measure. If the 
calibration film thickness is very different than the 
measurement film, the measurement will not be 
accurate.

Calibration Note

WARNING

The following icons are used to indicate important 

information which must be observed.

Denotes a 

prohibition

You MUST NOT do it.

Denotes a 

requirement

You MUST do it.

Improper use may result in personal
injury or damage to property.

CAUTION