13
CHECK/ADJUSTMENT PROCEDURE: Final Check
3
3.4
Final Check
○
Safety bevel check
Perform chamfering and have the user check the chamfered amount.
Depending on the user’s needs, change the setting on the SFB Setting screen.
See “Setting Parameters” of the operator’s manual for details.
○
Polish check (Type PLB only)
Polish a lens using the
∅
45 built-in data. If part of an edge remains unpolished, increase the setting
on the Differential screen. See “If part of an edge is not polished” of the operator’s manual for details.
Polish a lens using the
F
45 built-in data. If the whole edge is polished using the
∅
45 built-in data but
part of an edge remains unpolished using the
F
45 built-in data, the axis may be shifted in the polish-
ing process. In such a case, input the compensation value to the corresponding parameter setting
from “AXIS Constant PLA, BVL” to “AXIS Constant PC, FLT” on the AXIS Adjustment screen.
Have the user check the polished condition again. If the user indicates or questions something about
the polish, explain it is due to system performance.