background image

USB INTERFACE MANUAL

December 2010

Pages in total: 58

31001*RTZ002E/E

 LM-1000/1000P/1200

AUTO LENSMETER

Summary of Contents for LM-1000

Page 1: ...USB INTERFACE MANUAL December 2010 Pages in total 58 31001 RTZ002E E LM 1000 1000P 1200 AUTO LENSMETER ...

Page 2: ...ngo Tokyo Office Bunkyo Ku Tokyo 113 0033 Japan Telephone 81 3 5844 2641 Facsimile 81 3 5844 2642 NIDEK INCORPORATED 47651 Westinghouse Drive Fremont California 94539 U S A United States Agent Telephone 1 510 226 5700 Facsimile 1 510 226 5750 NIDEK S A Europarc 13 rue Auguste Perret 94042 Créteil France EU Authorized Representative Telephone 33 1 49 80 97 97 Facsimile 33 1 49 80 32 08 ...

Page 3: ...n 3 1 1 3 Installation of the USB driver 4 2 DATA FORMAT 5 2 1 Format of Entire Data 5 2 2 Format of each data 8 2 3 Example of Data Output 15 3 INSTALLATION OF USB DRIVER 37 3 1 Installation Procedure Windows 7 37 3 2 Installation Procedure Windows XP 42 3 3 Installation Procedure Windows 2000 47 3 4 Installation Procedure Windows 98 52 3 5 Installation Procedure Windows ME 54 ...

Page 4: ... 2 ...

Page 5: ...unication Among the parameters related to communication set the two parameters below The other parame ters related to communication have no influence on the data exported through the USB interface The underlined option shows the factory setting For the method of setting parameters see 2 13 Setting Parameters of the LM 1000 LM 1000P LM 1200 Operator s Manual Be sure to set the A Prt S and A Prt R L...

Page 6: ...e USB driver to receive the data measured by the LM 1000 LM 1000P or LM 1200 Install the provided USB driver attached to the USB cable option to the computer For the USB driver installation procedure refer to 3 INSTALLATION OF USB DRIVER page 37 Purchase a program used for data reception by yourself OFF The CR code is not added ON The CR code is added to the output character string OFF The measure...

Page 7: ...as been saved in the memory when the print button is pressed is transmitted During the measurement measured data that is in the memory when the print button is pressed is saved and transmitted 2 1 Format of Entire Data The format of the entire data to be transmitted is explained 1 SOH 01H This is referred to as communication control character and frequently used for data trans mission etc A single...

Page 8: ...8 8 EOT 04H This is a communication control character that represents the end of the character string In the case of measured data this character represents the end of data 9 CS1 CS4 10 CR This is a checksum for error detection The value of the checksum is the simple sum of SOH at the head to EOT in the data format except the CR code The hexadecimal num bers in the low order two bytes are represen...

Page 9: ... data is output in following order Right Left 4 In the prism layout the measured value is output for the PRISM value Measured lens mode Lens 3 Single vision Progressive power Contact lens Prism layout Single Right Left 1 Manufacturer model name Once at the beginning 2 SPH CYL AXIS 3 SE Spherical Equivalent value 4 ADD 1st 2nd addition power 2 5 PRISM 6 Progressive Length 1 7 Channel Width 1 8 PD 1...

Page 10: ... to the end when the CR Code parameter is set to ON The codes are added to the end of all the following types of data I D N I D E K L M 1 0 0 0 P ETB CR 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name LM 1000P I D N I D E K L M 1 0 0 0 ETB CR 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 17H 0DH Company name Model name LM 1000 I D N I D...

Page 11: ...f right data SE 10 00 D 0 1 0 0 0 0 0 0 0 0 0 ETB CR 20H 20H 2BH 30H 31H 2EH 30H 30H 2BH 30H 30H 2EH 30H 30H 30H 30H 30H 17H 0DH SPH 6 characters CYL 6 characters AXIS 3 characters R 1 1 2 5 0 9 7 5 0 9 0 ETB CR 20H 52H 2DH 31H 31H 2EH 32H 35H 2DH 30H 39H 2EH 37H 35H 30H 39H 30H 17H 0DH SPH 6 characters CYL 6 characters AXIS 3 characters L 0 0 0 0 0 1 5 0 1 8 0 ETB CR 20H 4CH 2BH 30H 30H 2EH 30H 3...

Page 12: ... 00 D 2 Example of the primary and secondary addition powers in the right data ADD 2 00 D ADD2 2 50 D 3 Example of the primary and secondary addition powers represented by spherical powers in the left data ADD 2 00 D ADD2 2 50 D NSPH 1 00 D NSPH2 0 50 D Identification code Single A A Space Right AR Left AL Identification code Single N N Space Right NR Left NL A 0 2 0 0 ETB CR 41H 20H 30H 32H 2EH 3...

Page 13: ...with no polarity sign and the leading digit 0 remaining 1 Example of single data when the PRISM parameter is set to BU D BI O PRISM IN 3 00Δ UP 2 50Δ 2 Example of right data when the PRISM parameter is set to BU D BI O PRISM OUT 1 25Δ DOWN 2 00Δ Identification code Single P P Space Right PR Left PL Identification code Single P B P Space B Space Right PR BR Left PL BL PRISM Four digits with a fixed...

Page 14: ...he progressive length is output by a two digit integer with the leading digit 0 remaining 1 Example of single data V Length 12 mm 2 Example of right data V Length 8 mm P L 0 6 5 0 ETB CR B L 0 7 0 ETB CR 50H 4CH 30H 36H 2EH 35H 30H 17H 0DH 42H 4CH 30H 37H 30H 17H 0DH PRISM 5 characters BASE 3 characters Identification code Single D D space Right DR Left DL D 1 2 ETB CR 44H 20H 31H 32H 17H 0DH V Le...

Page 15: ...data will not be included in the output unless PD has been measured for both sides The data is sent in the order of Total PD T PD Right PD R PD and Left PD L PD Measured data of the PD is output by fixed three digits including a decimal point If T PD is 100 0 mm or more 99 9 is output Sample of PD data T PD 64 0 mm R PD 31 5 mm L PD 32 5 mm Identification code Single W W space Right WR Left WL W 0...

Page 16: ... output unless prism has been measured for both sides The data is sent out in the order of Net Horizontal Prism and Net Vertical Prism The net prism measurements are always indicated by rectangular coordinates BASE IN OUT UP DOWN Measured data is output by four digits including a decimal point and the leading digit 0 remaining The base direction is shown with a letter put after the value just as i...

Page 17: ...ision lens measurement with the single state designated SPH 1 00 D CYL 0 00 D AXIS 0º No prism measurement performed CR code None Example outputs of 1 5 8 14 15 are common among LM 1000 LM 1000P and LM 1200 For the LM 1000 and LM 1200 replace the model name The following is an example of minimum data to be output SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 0 0 0 P ETB To ...

Page 18: ...H CYL and Axis data SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 0 0 0 P ETB To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H Company name Model name 0 1 0 0 0 0 0 0 0 0 0 ETB To be continued 20H 20H 2BH 30H 31H 2EH 30H 30H 2BH 30H 30H 2EH 30H 30H 30H 30H 30H 17H Single SPH Single CYL Single AXIS P 0 3 0 0 I ETB P 0 2 5 0 U 50H 20H 30H 33...

Page 19: ... added to the end of ETB and EOT SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 0 0 0 P ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name 0 2 0 0 0 0 5 0 0 6 0 ETB CR To be continued 20H 20H 2BH 30H 32H 2EH 30H 30H 2BH 30H 30H 2EH 35H 30H 30H 36H 30H 17H 0DH Single SPH Single CYL Single AXIS S 0 2 2 5 ETB C...

Page 20: ...y the identification code change from R to L SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 0 0 0 P ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name R 1 1 2 5 0 9 7 5 0 9 0 ETB CR To be continued 20H 52H 2DH 31H 31H 2EH 32H 35H 2DH 30H 39H 2EH 37H 35H 30H 39H 30H 17H 0DH Right SPH Right CYL Right AXIS P R ...

Page 21: ...X To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 0 0 0 P ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH Right SPH Right CYL Right AXIS A R 0 2 0 0 ETB CR 0 3 0 0 ETB CR To be continued 41H 52H 30H...

Page 22: ...01H 44H 4CH 4DH 02H I D N I D E K L M 1 2 0 0 ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 32H 30H 30H 17H 0DH Company name Model name R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH Right SPH Right CYL Right AXIS A R 0 2 0 0 ETB CR To be continued 41H 52H 30H 32H 2EH 30H 30H 17H 0DH Right ADD L 0 ...

Page 23: ... measured Net PRISM measurements will also be excluded from the output unless prism has been measured for both sides SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 2 0 0 ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 32H 30H 30H 17H 0DH Company name Model name R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2...

Page 24: ...FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH Right SPH Right CYL Right AXIS A R 0 2 0 0 ETB CR P R 0 2 5 0 I ETB CR To be continued 41H 52H 30H 32H 2EH 30H 30H 17H 0DH 50H 52H 30H 32H 2EH 35H 30H 49H 17H 0DH Right ADD Right PRISM1 P R 0 2 0 0 D ETB CR To ...

Page 25: ...m CR code is used PD data is added after R and L data Since the near portion inside amount of the left eye is not measured is output SOH D L M STX To be continued 01H 44H 4CH 4DH 02H I D N I D E K L M 1 2 0 0 ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 32H 30H 30H 17H 0DH Company name Model name R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 3...

Page 26: ...H 35H 17H 0DH Left ADD P D 6 4 0 3 1 5 3 2 5 ETB CR To be continued 50H 44H 36H 34H 2EH 30H 33H 31H 2EH 35H 33H 32H 2EH 35H 17H 0DH Total PD Right PD Left PD I S 0 1 5 ETB CR EOT 1 4 1 4 CR 49H 53H 2BH 30H 31H 2EH 35H 2AH 2AH 2AH 2AH 2AH 17H 0DH 04H 31H 34H 31H 34H 0DH Right INS Left INS Checksum ...

Page 27: ...racters in an ordinary lens measurement mode When the NEAR parameter is set to NEAR SPH both the addition powers ADD and near spherical data NSPH are output 1 5 Number of characters SOH D L M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I D N I D E K L M 1 0 0 0 P ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name 25 30 35 40 ...

Page 28: ...t SPH Left CYL Left AXIS 120 125 130 A L 0 2 2 5 ETB CR 0 3 5 0 ETB CR To be continued 41H 4CH 30H 32H 2EH 32H 35H 17H 0DH 30H 33H 2EH 35H 30H 17H 0DH Left ADD Left ADD2 135 140 145 N L 0 0 2 5 ETB CR 0 1 5 0 ETB CR To be continued 4EH 4CH 2BH 30H 30H 2EH 32H 35H 17H 0DH 2BH 30H 31H 2EH 35H 30H 17H 0DH Left NSPH Left NSPH2 150 155 160 165 P L 0 1 2 5 O ETB CR P L 0 2 0 0 U ETB CR To be continued 5...

Page 29: ... example of LM 1200 of the maximum data size 210 characters in ordinary lens measurement mode When the NEAR parameter is set to NEAR SPH both the addition powers ADD and near spherical data NSPH are output When the Net PRISM parameter is set to ON differences between right and left prism values are included in the output as the Net PRISM value 1 5 Number of characters SOH D L M STX To be continued...

Page 30: ...ed 41H 4CH 30H 32H 2EH 32H 35H 17H 0DH 30H 33H 2EH 35H 30H 17H 0DH Left ADD Left ADD2 135 140 145 N L 0 0 2 5 ETB CR 0 1 5 0 ETB CR To be continued 4EH 4CH 2BH 30H 30H 2EH 32H 35H 17H 0DH 2BH 30H 31H 2EH 35H 30H 17H 0DH Left NSPH Left NSPH2 150 155 160 165 P L 0 1 2 5 O ETB CR P L 0 2 0 0 U ETB CR To be continued 50H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH Left ...

Page 31: ...urement mode When the NEAR parameter is set to NEAR SPH both the addition powers ADD and near spherical data NSPH are output 1 5 Number of characters SOH D L M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I D N I D E K L M 1 0 0 0 P ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name 25 30 35 40 R 0 1 2 5 0 0 7 5 1 2 0 ETB CR T...

Page 32: ...eft AXIS 105 110 115 A L 0 2 2 5 ETB CR N L 0 0 2 5 ETB CR To be continued 41H 4CH 30H 32H 2EH 32H 35H 17H 0DH 4EH 4CH 2BH 30H 30H 2EH 32H 35H 17H 0DH Left ADD Left NSPH 120 125 130 135 P L 0 1 2 5 O ETB CR P L 0 2 0 0 U ETB CR To be continued 50H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH Left PRISM1 Left PRISM2 140 145 EOT 1 A 7 0 CR 04H 31H 41H 37H 30H 0DH ...

Page 33: ...m Position 18 mm LPD 32 5 mm LINS 2 0 mm Total PD 64 0 mm CR code Added This is an example of the LM 1200 of the maximum data size 204 characters in the progressive power lens measurement mode When the NEAR parameter is set to NEAR SPH both the addition powers ADD and near spherical data NSPH are output When the function of the progressive length measurement and or the channel width measurement is...

Page 34: ...8H 30H 17H 0DH Left SPH Left CYL Left AXIS 115 120 125 130 A L 0 2 2 5 ETB CR N L 0 0 2 5 ETB CR To be continued 41H 4CH 30H 32H 2EH 32H 35H 17H 0DH 4EH 4CH 2BH 30H 30H 2EH 32H 35H 17H 0DH Left ADD Left NSPH 135 140 145 150 P L 0 1 2 5 O ETB CR P L 0 2 0 0 U ETB CR To be continued 50H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH Left PRISM1 Left PRISM2 155 160 165 D ...

Page 35: ... M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I D N I D E K L M 1 0 0 0 P ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name 25 30 35 40 R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH Right SPH Right CYL Right AXIS 45 50 55 60 S R 0 1 5 0 ETB CR P R ...

Page 36: ...P L 0 1 2 5 O ETB CR To be continued 53H 4CH 2DH 30H 32H 2EH 32H 35H 17H 0DH 50H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH Left SE Left PRISM1 115 120 125 127 P L 0 2 0 0 U ETB CR EOT 1 7 4 3 CR 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH 04H 31H 37H 34H 33H 0DH Left PRISM2 Checksum ...

Page 37: ... measured prism power 1 5 Number of characters SOH D L M STX To be continued 01H 44H 4CH 4DH 02H 10 15 20 I D N I D E K L M 1 0 0 0 P ETB CR To be continued 49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 30H 30H 30H 50H 17H 0DH Company name Model name 25 30 35 40 R 0 1 2 5 0 0 7 5 1 2 0 ETB CR To be continued 20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH Right SPH Ri...

Page 38: ...H 4CH 30H 31H 2EH 32H 35H 4FH 17H 0DH 50H 4CH 30H 32H 2EH 30H 30H 55H 17H 0DH Left PRISM1 Left PRISM2 100 105 110 115 P D 6 4 0 3 1 5 3 2 5 ETB CR To be continued 50H 44H 36H 34H 2EH 30H 33H 31H 2EH 35H 33H 32H 2EH 35H 17H 0DH Total PD Right PD Left PD 116 120 121 EOT 1 6 8 E CR 04H 31H 36H 38H 45H 0DH Checksum ...

Page 39: ...uter running Windows 7 Log in as an administrator Some procedures might not be displayed due to security setting 1 Connect the LM 1000 LM 1000P or LM 1200 and computer with the USB cable 2 Turn on the LM 1000 LM 1000P or LM 1200 The operating system recognizes unknown devices 3 Open the Property in My Computer 4 Open the Device Manager The new dialog is displayed 5 Open the added Lensmeter USB Dev...

Page 40: ...mputer for driver software The new dialog opens 8 Press the Browse button The Browse for Folder dialog opens 9 Designate the location of the FTDI BUS INF file and press the OK button Ordinarily designate the E001 Windows7 folder in the distributed media The Browse for Folder dialog closes 10 Press the Next button The Windows 7 Security dialog opens ...

Page 41: ...RIVER Installation Procedure Windows 7 3 11 Choose the Install this driver software anyway The new dialog is displayed 12 Press the Close button 13 Press the Close button 14 Open the USB Serial Port The new dialog is displayed ...

Page 42: ...Browse my computer for driver software 17 Press the Browse button The Browse for Folder dialog opens 18 Designate the location of the FTDIPORT INF file and press the OK button Ordinarily designate the E001 Windows7 folder in the distributed media The Browse for Folder dialog closes 19 Press the Next button The Windows Security dialog opens ...

Page 43: ...ocedure Windows 7 3 20 Choose the Install this driver software anyway The new dialog is displayed 21 Press the Close button 22 Press the Close button 23 Turn off the LM 1000 LM 1000P or LM 1200 The installation of the USB driver has been completed ...

Page 44: ...Device The Welcome to the Found New Hardware Wizard dialog is displayed 3 Press the No not this time radio button and press the Next button The new dialog is displayed 4 Choose the Install from a list or specific location Advanced radio button and press the Next button The Please choose your search and installation options dialog is displayed 5 Choose the Search for the best driver in these locati...

Page 45: ...older in the distributed media The browse for Folder dialog opens 7 Press the Next button The Hardware Installation dialog opens 8 Press the Continue Anyway button The Completing the Found New Hardware Wizard dialog is dis played 9 Press the Finish button 10 The plug play operating system subsequently recognizes the USB Serial Port The Welcome to the Found New Hardware Wizard dialog is displayed ...

Page 46: ...Select the Install from a list or specific location Advanced radio button and click the Next button The Please choose your search and installation options dialog is displayed 13 Choose the Search for the best driver in these locations radio button then choose the Include this location in the search check box and press the Browse button The Browse For Folder dialog opens ...

Page 47: ...T INF file and press the OK button Ordinarily designate the E001 Windows2000 2003 XP Vista folder in the distributed media The Browse For Folder dialog closes 15 Press the Next button The Hardware Installation dialog opens 16 Press the Continue Anyway button The Found New Hardware Wizard dialog is displayed ...

Page 48: ... Installation Procedure Windows XP 17 Press the Finish button 18 After using the Safety Remove Hardware function remove the USB cable 19 Turn off the LM 1000 LM 1000P or LM 1200 The installation of the USB driver has been completed ...

Page 49: ...e plug play operating system recog nizes the LM 1000 1200 USB Device The Welcome to the Found New Hard ware Wizard dialog is displayed 3 Click the Next button The Install Hardware Device Drivers dia log is displayed 4 Choose the Search for a suitable driver for my device recommended radio but ton and press the Next button The Locate Driver Files dialog is dis played 5 Only check the Specify a loca...

Page 50: ...Look in box Ordinarily designate the E001 Windows2000 2003 XP Vista folder in the distributed media 8 Select the FTDIBUS INF file and press the Open button The dialog for designating a file is activated 9 Press the OK button The Driver Files Search Results dialog is activated 10 Press the Next button The Completing the Found New Hardware Wizard dialog is dis played ...

Page 51: ...ntly recognizes the USB Serial Port The Welcome to the Found New Hard ware Wizard dialog is displayed 12 Press the Next button The Install Hardware Device Drivers dia log is displayed 13 Choose the Search for a suitable driver for my device recommended radio but ton and press the Next button The Locate Driver Files dialog is dis played ...

Page 52: ...the Browse button The Locate File dialog is displayed 16 Designate the folder including the FTDI PORT INF file in the Look in box Ordinarily designate the E001 Windows2000 2003 XP Vista folder in the distributed media 17 Select the FTDIPORT INF file and press the Open button The dialog for designating a file is activated 18 Press the OK button The Driver Files Search Results dialog is activated ...

Page 53: ...the Next button The Completing the Found New Hard ware Wizard dialog is displayed 20 Press the Finish button 21 After using the Safely Remove Hardware function remove the USB cable 22 Turn off the LM 1000 LM 1000P or LM 1200 The installation of the USB driver has been completed ...

Page 54: ...the LM 1000 LM 1000P or LM 1200 The plug play operating system recognizes the LM 1000 1200 USB Device The Add New Hardware Wizard dialog is displayed 3 Press the Next button The new dialog is displayed 4 Choose the Search for the best driver for your device Recommended radio but ton and press the Next button The new dialog is displayed 5 Only check the Specify a location check box and press the Br...

Page 55: ...distributed media 7 Press the OK button The Browse for Folder dialog closes 8 Press the Next button The new dialog is displayed 9 Press the Next button The new dialog is displayed 10 Press the Finish button 11 After using the Safely Remove Hardware function remove the USB cable 12 Turn off the LM 1000 LM 1000P or LM 1200 The installation of the USB driver has been completed ...

Page 56: ...e The Add New Hardware Wizard dialog is displayed 3 Choose the Specify the location of the driver Advanced radio button and press the Next button The new dialog is displayed 4 Choose the Search for the best driver for your device Recommended radio but ton only check the Specify a location check box and press the Browse button The Browse for Folder dialog opens 5 Select the folder including the FTD...

Page 57: ... button The new dialog is displayed 8 Press the Next button The new dialog is displayed 9 Press the Finish button 10 After using the Safely Remove Hardware function remove the USB cable 11 Turn off the LM 1000 LM 1000P or LM 1200 The installation of the USB driver has been completed ...

Page 58: ...56 INSTALLATION OF USB DRIVER Installation Procedure Windows ME ...