4 - 11
MCP8**RDA001A/E
4.8 The Chart does not Appear Properly When Looking at Polarized
Chart through the Polarizing Glasses.
4.9 The Beep Volume is Inappropriate.
Install the main body on a level surface.
Is the main body tilted?
Is the projected chart tilted?
Is the refractor used as the check instrument for
test subjects?
Is the trial lens used as the check instrument for
test subjects?
When the polarizing glasses are used as the
check tool for subjects and the polarizing glasses
are reversed left and right, is the appearance of
the polarized chart changed?
Replace the chart ASSY. (See 6..1.6.2.)
Adjust the initial position. (See 7.3.)
Replace the chart ASSY. (See 6.1.6.2.)
Replace the polarizing glasses.
Does the polarized chart disappear when the trial
lens is moved?
Yes
No
Yes
No
Yes
No
Yes
No
Yes
No
Replace the chart ASSY. (See 6.1.6.2.)
Completion
Yes
No
Adjust the beep volume.
(Refer to the operator
㵭
s manual.)
Is the symptom improved?
Replace the main board. (34801-BA01).
(See 6.2.)
Completion
Yes
No
Summary of Contents for CP-770
Page 1: ...CHART PROJECTOR Model CP 770 SERVICE MANUAL September 2007 Pages in total 64 MCP8 RDA001A E...
Page 6: ...MCP8 RDA001A 1 2 This page is intentionally left blank...
Page 10: ...3 2 MCP8 RDA001A E This page is intentionally left blank...
Page 22: ...4 12 MCP8 RDA001A E This page is intentionally left blank...
Page 34: ...6 10 MCP8 RDA001A E This page is intentionally left blank...
Page 51: ...8 1 MCP8 RDA001A E 8 INFORMATION 8 1 Wiring Diagram...
Page 52: ...8 2 MCP8 RDA001A E 8 2 Connector Cable 34801 EA01 34801 EA02 34801 EA03 34801 EA04...
Page 53: ...8 3 MCP8 RDA001A E 34801 EA06 34801 EA07 34801 EA08...
Page 54: ...8 4 MCP8 RDA001A E 34801 E004...
Page 57: ...8 7 MCP8 RDA001A E 8 4 Label...