FP-AIO-600 and cFP-AIO-600

This section describes the isolation of the [c]FP-AIO-600 and its 
compliance with international safety standards. The field wiring 
connections are isolated from the backplane and the inter-module 
communication bus. The isolation barriers in the module provide 
250 V

rms

 Measurement Category II continuous isolation, verified 

by 2,300 V

rms

, 5 s dielectric withstand test. The [c]FP-AIO-600 

provides 

double insulation

 (compliant with IEC 61010-1) for 

working voltages of 250 V

rms

1

. Safety standards (such as those 

published by UL and IEC) require the use of double insulation 
between hazardous voltages and any human-accessible parts or 
circuits.

Never

 try to use any isolation product between human-accessible 

parts (such as DIN rails or monitoring stations) and circuits that 
can be at hazardous potentials under normal conditions, unless the 
product is specifically designed for such an application, as is the 
[c]FP-AIO-600.

Even though the [c]FP-AIO-600 is designed to handle applications 
with hazardous potentials, follow these guidelines to ensure a safe 
total system:

There is no isolation between channels on the [c]FP-AIO-600. 
If a hazardous voltage is present on any channel, all channels 
are considered hazardous. Make sure that all other devices and 
circuits connected to the module are properly insulated from 
human contact.

Do 

not

 share the external supply voltages (the V and C 

terminals) with other devices (including other FieldPoint 
devices), unless those devices are isolated from human contact.

For Compact FieldPoint, you 

must

 connect the protective earth 

(PE) ground terminal on the cFP-BP-

x

 backplane to the system 

safety ground. The backplane PE ground terminal has the 
following symbol stamped beside it: 

. Connect the 

backplane PE ground terminal to the system safety ground 
using 14 AWG (1.6 mm) wire with a ring lug. Use the 5/16 in. 
panhead screw shipped with the backplane to secure the ring 
lug to the backplane PE ground terminal.

As with any hazardous voltage wiring, make sure that all 
wiring and connections meet applicable electrical codes and 
commonsense practices. Mount terminal bases and backplanes 

1  

Working voltage

 is defined as the signal voltage plus the common-mode voltage. 

Common-mode voltage

 is the voltage of the module with respect to ground.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for FieldPoint cFP-AIO-600

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...IO 600 is a FieldPoint analog input and output module with the following features Four analog voltage or current input channels with 11 input ranges up to 36 V or 24 mA with overranging Four analog cu...

Page 3: ...with the guide rails on the terminal base 3 Press firmly to seat the FP AIO 600 on the terminal base When the module is firmly seated the terminal base latch locks it into place Figure 1 Installing t...

Page 4: ...s firmly to seat the cFP AIO 600 on the backplane 3 Using a number 2 Phillips screwdriver with a shank of at least 64 mm 2 5 in length tighten the captive screws to 1 1 N m 10 lb in of torque The nylo...

Page 5: ...l power supply that provides at least 125 mA for the output channels The power supply must provide more current if you use it to power transducers or field devices Connect the external power supply to...

Page 6: ...All four input channels share a common ground reference that is isolated from other modules in the FieldPoint system Table 1 Input Terminal Assignments Channel Inputs VIN IIN 1 VSUP 2 COM In 0 1 2 17...

Page 7: ...30 V FieldPoint software accounts for the 20 overranging feature and shows the ranges accordingly Figure 5 shows how to connect a voltage source without an external power supply to one input channel o...

Page 8: ...ture and shows the ranges accordingly Figure 7 shows how to connect a current source without an external power supply to one input channel of the c FP AIO 600 Figure 7 Current Input without an Externa...

Page 9: ...r field devices with span errors of up to 20 of full scale Also the overranging feature helps prevent errors caused by clipping Connecting the c FP AIO 600 to a Load The c FP AIO 600 has four single e...

Page 10: ...s a monitoring circuit that compares the actual output current to the specified output current If the c FP AIO 600 cannot source the specified output current for one or more channels the monitoring ci...

Page 11: ...lights and the c FP AIO 600 informs the network module of its presence When the network module recognizes the c FP AIO 600 it sends initial configuration information to the c FP AIO 600 After the c FP...

Page 12: ...total system There is no isolation between channels on the c FP AIO 600 If a hazardous voltage is present on any channel all channels are considered hazardous Make sure that all other devices and circ...

Page 13: ...4 hazardous locations Class I Zone 2 AEx nA IIC T4 and Ex nA IIC T4 hazardous locations and nonhazardous locations only Follow these guidelines if you are installing the cFP AIO 600 in a potentially e...

Page 14: ...are maximums for the range 40 to 70 C unless otherwise noted Specifications are subject to change without notice Input and Output Characteristics Number of input channels 4 Number of output channels 4...

Page 15: ...7 12 V 0 04 0 10 0 19 0 12 18 V 0 04 0 09 0 19 0 11 36 V 0 04 0 08 0 19 0 09 Current Input Range with Overranging Accuracy at 15 to 35 C of Reading of Full Scale Accuracy at 40 to 70 C of Reading of F...

Page 16: ...y by current output range Dynamic Characteristics Hardware update rate 1 700 updates s Slew rate 0 4 mA s Physical Characteristics Indicators Green POWER and READY indicators and 4 red output indicato...

Page 17: ...Humidity 10 to 90 RH noncondensing Maximum altitude 2 000 m at higher altitudes the isolation voltage ratings must be lowered Pollution Degree 2 Shock and Vibration These specifications apply only to...

Page 18: ...ement control and laboratory use EN 61326 EMC requirements Industrial Immunity EN 55011 Emissions Group 1 Class A CE C Tick ICES and FCC Part 15 Emissions Class A Note For EMC compliance operate this...

Page 19: ...ronment weee htm Mechanical Dimensions Figure 10 shows the mechanical dimensions of the FP AIO 600 installed on a terminal base If you are using the cFP AIO 600 refer to your Compact FieldPoint contro...

Page 20: ...43 0 662 45 79 90 0 Belgium 32 0 2 757 00 20 Brazil 55 11 3262 3599 Canada 800 433 3488 China 86 21 6555 7838 Czech Republic 420 224 235 774 Denmark 45 45 76 26 00 Finland 385 0 9 725 725 11 France 3...

Page 21: ...more information about National Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instr...

Page 22: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Reviews: