indicates specific error conditions by flashing the Status LED a certain number of times every
few seconds, as shown in the following table.
Table 2. Status LED Indications
Number of Flashes
Every Few Seconds
Indication
2
The device has detected an error in its software. This usually
occurs when an attempt to upgrade the software is interrupted.
Reinstall software on the device.
3
The device is in safe mode.
4
The software has crashed twice without rebooting or cycling
power between crashes. This usually occurs when the device runs
out of memory. Review your LabVIEW Real-Time VI and check
the memory usage. Modify the VI as necessary to solve the
memory usage issue.
Continuously flashing or
solid
The device has detected an unrecoverable error. Contact NI.
LEDs 0-3
You can use LEDs 0-3 to help debug your application or easily retrieve application status.
Logic TRUE turns an LED on and logic FALSE turns an LED off.
Physical Dimensions
The following figures describe the physical dimensions of the NI ELVIS RIO CM enclosure
and its features.
Figure 8. NI ELVIS RIO CM Dimensions, Top View
2X 11.2 mm
(0.44 in.)
247.5 mm
(9.74 in.)
161.0 mm
(2.40 in.)
272.3 mm
(10.72 in.)
48.0 mm
(3.76 in.)
2X
Ø
11.2 mm
(0.44 in.)
8
|
ni.com
|
NI ELVIS RIO CM User Manual