NI PXIe-5624R Calibration Procedure
|
© National Instruments
|
13
4.
Connect the 6 dB attenuator SMA (m) connector directly to the NI 5624R IF IN front panel
connector.
The following figure illustrates the complete hardware setup.
Figure 4.
Absolute Amplitude Accuracy Verification Cabling Diagram
1
Power Sensor
2
SMA (m)-to-N (f) Adapter
3
Power Splitter
4
3.5 mm (m)-to-3.5 mm (m) Adapter
5
6 dB Attenuator
6
NI 5624R
7
SMA (m)-to-SMA (m) Cable
8
Signal Generator
ACCESS
ACTIVE
4
5
7
3
2
1
6
8