6-6
WF1946B
6.2 Suspected failure
In case of abnormality
If an abnormality is suspected, check as indicated in the following table. If normal operation cannot be
returned, contact the dealer.
Symptom
Possible
causes
Correction
Power source not within
specified range
Check the rear panel power supply rating.
Use the equipment at single phase AC, at the rated
voltage
±
10 %, 50 to 60 Hz (48 to 62 Hz).
Power supply fuse open
Replace the power supply fuse.
(Be sure to use the correctly rated fuse.)
No power on
External noise
Install the equipment in a site with favorable
conditions.
Remote mode enabled
Press the LOCAL key to set the local mode.
Panel inoperative
Keys or Modify dial
defective
Contact service.
Ambient temperature and
humidity outside specified
range
Use the equipment under the specified
environmental conditions.
Inadequate warm up
Allow the equipment to warm up for at least 30
minutes after power on.
DC offset applied
Set DC offset to 0 V.
Set for user units
Select standard units
Output abnormal
LOAD function being used
Set for OPEN.
Address or USB ID
different from that in the
program
Set the address and USB ID to match those in the
program.
Cannot be set via
remote control
Address or USB ID identical
to that of another device
Set the address and USB ID so that they do not
match those of other devices.
Setting have not been
initialized (PRESET)
The operation manual presumes the settings have
been initialized.
Does not operate
according to
operation manual
Operated channel reversed
Check if CH1 or CH2.
Summary of Contents for WF1946B
Page 2: ......
Page 3: ...DA00012506 002 WF1946B MULTIFUNCTION SYNTHESIZER Instruction Manual ...
Page 4: ......
Page 8: ......
Page 28: ......
Page 66: ......
Page 107: ...4 2 Sweep WF1946B 4 41 ...
Page 164: ......
Page 180: ......
Page 198: ......
Page 200: ......
Page 202: ......
Page 203: ......