7.3 Performance tests
WF1946B
7-9
Time difference between channels
Connections:
CH1 Function Out to universal counter input 1 (50
Ω
terminated)
CH2 Function Out to universal counter input 2 (50
Ω
terminated)
Use the same length and same type coaxial cables.
Settings:
After setting initialize, set channel mode 2PHASE, amplitude 20 Vp-p/open (CH1 and
CH2), CH2 phase 180 deg, and frequency and waveform according to the following
table.
Measurement:
Set timer mode between counter inputs 1 and 2, and measure the time difference
between CH1 and CH2. Set the counter trigger level to 0 V and the trigger polarity to
rising for both inputs 1 and 2. Use the average value, since dispersion occurs in the
measured values.
Determination:
The ranges indicated in the following table are normal.
Waveform Frequency
Rating
range
10 MHz
40 to 60 ns
(duty 50 % fixed)
10 MHz
40 to 60 ns
500 kHz
990 to 1010 ns