5.6
Notes for P rogramming
DF1906
5-19
5.6 Notes for Programming
■ Remote
There is no remote state for this equipment. The panel operation is possible while the equipment is
controlled by a PC, so make sure that you do not accidentally change the settings by unintentional panel
operation.
■ Arbitrary Waveform Data
Arbitrary waveform data is a binary data block, so transfer it separately from the proceeding command
part of the ASCII program message. Be sure to specify 4096 words (8192 bytes) when transferring the
arbitrary waveform data. If the data exceeds or falls short of that size, an error occurs or a correct
waveform cannot output.
■ Note for Sending Commands from USB Interface
When sending a command, LF (0AH) must be added as a program message terminator to the end of the
string to be sent. If a command is sent without adding LF, it will not work correctly.