5.3
Explanation of Commands
DF1906
5-13
・
:TRIGger[:SEQuence]:SLOPe {POSitive |NEGative}
:TRIGger[:SEQue nce]:SLOPe ?
Explanation: Sets/queries the slope where the trigger input is validated.
Parameter:
POSitive (Rise)
NEGative (Fall)
Response: Returns POS or NEG.
ex)
TRIG:SLOP?
POS
・
[:SOURce]:BURSt:GATE:POLarity {NORMal|INVerted}
[:SOURce]:BURSt:GATE:POLar ity?
Explanation: Sets/queries the level where the gate input is validated.
Parameter:
NORMal(HIGH)
INVerted(LOW)
Response: Returns NORM or INV.
ex)
BURS:GATE:POL?
NORM
・
:OUTPut[:STATe] {OFF|ON}
:OUTPut[:STATe ]?
Explanation: Sets/queries the output on/off.
Parameter:
OFF (Output off)
ON (Output on)
Response: Returns 0 (OFF) or 1 (ON).
ex)
OUTP?
1
・
:MEMory:STATe :DELete
Explanation: Initializes the selected setting memory.