21
•
Patients with pacemakers or physicians should exercise caution when using it.
2 > AIR SUPPLY PRESSURE
Set the pressure at the back end of the handpiece to 206~231 KPa (2.1~2.35 kgfcm2).
If a net use pressure gauge is available, set the air pressure to 241~275 KPa (2.45~2.8 kgf/cm2) on the drive air pressure gauge on the
unit.
Under the above recommended operating pressure, the air consumption is< 1.5 L/min.
CAUTION
•
Do not exceed the recommended air pressure. This is important to avoid higher speed which could cause excessive bearing failure.
•
To prevent the life of the unit, it is essential to use dry, purified compressed air and to maintain air and water quality by regular
maintenance of the compressor and filtration systems. The use of unfiltered hard water will cause early blockage of tubing, connec-
tors and spray cones.
•
The life cycle of this product should be in accordance with the requirements of the relative local laws.
3 > CAUTIONS ON BURRS
•
ISO standard shank diameter is 1.59mm~1.60mm; standard length is 21mm~23mm.
Summary of Contents for Rose 202-M4 2SH
Page 68: ...68 ...
Page 73: ...73 2 PARÁMETROS TÉCNICOS 1 Motor de aire de agua interno 2 Pieza recta ...
Page 86: ...86 2 Straight handpiece 3 Contraangle handpiece ...
Page 98: ...98 2 PARAMÈTRES TECHNIQUES 1 Moteur interne eau air 2 Pièce droite ...
Page 111: ...111 2 PARÂMETROS TÉCNICOS 1 Motor de ar aquático interno 2 Peça recta ...
Page 124: ...124 2 PARAMETRI TECNICI 1 Motore ad aria con acqua interna 2 Pezzo diritto ...