Command Set
63
SL
set left travel limit
USAGE
IMM
♦
MIP
SYNTAX
xxSLnn or xxSL?
PARAMETERS
Description
xx
[ int ]
-
Controller Number
nn
[ int ]
-
Negative Software limit
Range
xx
-
0
to
255
nn
-
-60000
to 1200000
Units
xx
-
None
nn
-
Micro-steps
Defaults
xx
Missing:
None
out of range:
None
nn
missing:
Error 37, Parameter Missing
out of range:
Error 7, Parameter Out of Range
Error 230, Command not allowed
during motion
DESCRIPTION
This command will set the value of the negative software limits.
NOTE:
The value of Negative software limit should be less than
the value of the Positive Software limit.
RETURNS
If the “?” sign takes the place of
nn
value, this command reports
the value of the Negative Software limit.
REL.
COMMANDS
SR
-
Positive Software Limit
PA
Absolute Move
PR
Relative Move
JA
-
Jog
EXAMPLE
72SL-100
|
Sets the value of Negative software limit
to –100 for controller 72
725SL?
|
Reads back the value of the Negative
Software Limit from controller 72
Summary of Contents for NewStep NSA12
Page 1: ...Actuator and Motion Controller NewStep User s Manual TM...
Page 20: ...20 Getting Started This page is intentionally left blank...
Page 25: ...Operating NewStep 25 4 1 3 Single Actuator no Computer Interface minimal configuration...
Page 26: ...26 Operating NewStep 4 1 4 Single Actuator with Computer Interface...
Page 27: ...Operating NewStep 27 4 1 5 Two Actuators no Computer Interface...
Page 28: ...28 Operating NewStep 4 1 6 Two Actuators with Computer Interface...
Page 29: ...Operating NewStep 29 4 1 7 Multiple Actuators with Computer Interface...
Page 32: ...32 Operating NewStep This page is intentionally left blank...