Model ESA-CXA Electrostrictive Actuator Controller
Actuator Compatibility
This unit is intended for use only with Newport AD Series
electrostrictive actuators.
Caution
Clamp-type micrometer/actuator mounts are
capable of very high clamping forces. Do not
over-clamp the actuators! Tighten the clamp
only as much as necessary to firmly grip the
actuator. Performance degradation and/or
damage can occur if the clamps are
overtightened.
Front Panel of ESA-CXA Controller
ON
OFF
AXIS ADJUSTMENT
X Y Z
X
Y
Z
AXIS
DISPLAY
■
Drive Controller
■
Model ESA-CXA
I
O
3
1
4
2
Figure 1: Front panel of the ESA-CXA controller
[Refer to numbers in Figure 1]
1. On/Off switch
Power switch to turn the power of the ESA-CXA on and
off. Be sure the controller is set to the “OFF ” position
when connecting actuators.
Caution
Do not insert or remove actuator connectors
while power is on. Amplifier damage may result.
2. Voltage adjust knob
The DC voltage can be adjusted with the 10-turn
potentiometer and is limited from approximately 10 to
160 volts. Since voltage may be present at the output
connector on the controller, do not probe this area with
any conductive material.
3. Voltage display
The LED readout display shows the voltage output to an
actuator which is set by the voltage adjust knob (2) for
X, Y, or Z axis.
Operation of ESA-CXA
Electrostrictive Actuator Controller
Read and understand all instructions before operating.
Safety Terms and Symbols
The Following safety terms and symbols are uses in this
manual and/or on the controller.
1. Terms
The
Warning
heading explains dangers that could result in
personal injury or death.
The
Caution
heading explains hazards that could damage the
controller.
2. Symbols
Refer to the documentation for proper operating
instructions.
Protective earth ground connection
Fuse symbol
On (power)
Off (power)
Warning
The following precautions should be observed
at all times:
•
Care must be taken as high voltages are present
inside the controller and at the actuators. Do not
operate the unit with damaged cables or with the
controller opened.
•
Do not operate the unit with a damaged power cord
or if the controller has been dropped or damaged.
Contact Newport immediately and arrange to return
it for service.
Warning
To avoid the danger of potential electrical shock
to the user:
This instrument is intended for use by qualified
personnel who recognize shock hazards and
are familiar with safety precautions to avoid
possible injury.
•
Keep all fluids away from the controller and
actuators.
•
Do not insert metal (conductive) objects into the
rear panel connectors or vent holes.
•
Do not remove the cover!
There are no user
serviceable components inside the controller.
•
Refer all service requirements to the factory or
Newport authorized service facility.