
62
WARNING
To reduce the risk of electric shock or damage to the instrument,
turn the power switch off and disconnect the power cord before
replacing a fuse.
If a fuse blows:
1.
Disconnect the power cord from the controller
2.
Pry open tab at top of power receptacle and open panel to access fuses.
(See Figure 22)
3.
Replace the fuse(s).
Use only 5mm x 20mm fuses as indicated below
Fuse Replacement
Line Voltage
1.6 Amp, (T), 250V
100/120VAC
0.8 Amp, (T), 250V
220/240VAC
4.
Reconnect the power cord and turn on the instrument.
5.
If the problem persists, contact Newport Corporation for service.
8.3
Obtaining Service
The Model 350B Temperature Controller contains no user serviceable parts. To obtain
information regarding service, contact Newport Corporation or a Newport representative.
Please have the following information available:
1.
Instrument model number (on the rear panel)
2.
Instrument serial number (on rear panel)
3.
Description of the problem.
If the instrument is to be returned to Newport Corporation, a Return Number will be
issued, which should be referenced in the shipping documents. Please fill out a copy of
the service form, located on the following page, and have the information ready when
contacting Newport Corporation. Return the completed service form with the instrument.
Summary of Contents for 350B
Page 2: ...Temperature Controllers User s Manual Model 350B ...
Page 7: ...vi This page is intentionally left blank ...
Page 11: ...x This page is intentionally left blank ...
Page 25: ...24 Getting Started This page is intentionally left blank ...
Page 31: ...30 System Operation This page is intentionally left blank ...
Page 61: ...60 ...