40
Computer Interfacing
TEC:Ite
Description
TEC I
TE
set point command.
Syntax
TEC:Ite
set point
Remarks
The
TEC:Ite
command sets the TEC control current set point.
Argument
Description
set point
set point in Amps
See Also
TEC:Ite?, TEC:LIMit:Ite, TEC:SET:Ite?
TEC:Ite?
Description
TEC measured I
TE
query.
Syntax
TEC:Ite?
Remarks
The
TEC:Ite?
query returns the value of the measured TEC output current.
Response
Description
measured output
Current in Amps
The TEC current is constantly measured and updated, regardless of the TEC mode of operation.
This measurement is updated approximately once every 225 milliseconds.
See Also
TEC:Ite
TEC:LIMit:Ite
Description
TEC ITE current limit command
Syntax
TEC:LIMit:Ite
limit
Remarks
The
TEC:LIMit:Ite
command sets the TEC ITE current limit value.
Argument
Description
limit
Limit in Amps
The factory default current limit is 50% of the maximum current.
See Also
TEC:Ite