Newport 1815-C Operator'S Manual Download Page 25

12

3.3.4  Ambient and Stray Light

Ambient and stray light striking the detector will be measured by the Model
1815-C, and should be considered when making careful measurements. Ambi-
ent light can be distinguished from dark current (or the detector/meter noise
floor) by either turning off or blocking the source 

and

 covering the detector

face with opaque material such as a piece of black metal.  Using the human
hand to cover the detector is not advised because it both emits a significant
amount of infrared radiation, and because it radiates a temperature signifi-
cantly different from ambient.  With the detector covered, a reading of the
dark current may be made.  Next, remove the material which is covering the
detector and take another reading.  The difference is the ambient light level.

The effects of ambient light are greatly reduce when using a fiber-
connectorized signal input to the detector.  If free-space beam measurements
are desired, using an attenuator (

Low-Power

 detectors have an OD3 attenua-

tor included) will reduce stray light and often improve the source signal to
ambient signal noise level.  Wavelength-specific filters, such as optical cutoff,
bandpass, or spike filters can also be used if the signal wavelength spectrum
permits.  Other techniques to reduce stray light include using apertures,
placing the detector in a box or other housing to shield the surface from light
(or air currents when using Newport’s 

High-Power

 disk thermopile detectors)

which is not coming from the source, and turning off room and other lights.

Note

Changes in ambient light levels can occur from such factors as
turning room lights on or off, or by moving people or equipment.
Remember, if you can see your detector element, then your detector
can see the light bouncing off your shirt!

Basic measurement techniques for using the Model 1815-C are covered in the
following sections.  Also included are methods of background correction and
common measurement errors.  In general the absolute measurement accuracy
is limited by the accuracy of the detector calibration and environmental
factors affecting the detector.  See the appropriate detector manual for
specific information on a particular detector model.

3.4.1  Making a Power Measurement

The following process describes the procedure for making a basic optical
power measurement while properly removing influence of ambient light and
other drift effects.

1. Set up the meter calibration factor to reflect the detector at wavelength

of light you will be making your power measurement at.  See Sections
2.4.4.1 and 2.4.4.2.

2. Cover or otherwise block the source that you will be measuring and

adjust the Range knob until the  ambient signal is displayed to three
significant figures.

3. Use the ZERO knob to remove the ambient signal by zeroing the display.

4. Uncover the source and adjust the Range knob until the source power is

displayed to three significant figures of accuracy.

5. Record the display value and the appropriate units from the Range

knobs position. This reading is your optical power.

Performing Basic
Measurements

3.4

Summary of Contents for 1815-C

Page 1: ...Model 1815 C Optical Power Meter OPERATOR MANUAL...

Page 2: ......

Page 3: ...anty This warranty does not apply to defects resulting from modification or misuse of any product or part This warranty also does not apply to fuses batteries or damage from battery leakage This warra...

Page 4: ...r EN55011 Group 1 Class A EN50082 1 IMMUNITY Electrostatic Discharge per IEC 1000 4 2 severity level 3 Radiated Emission Immunity per IEC 1000 4 3 severity level 2 Fast Burst Transients per IEC 1000 4...

Page 5: ...Using the System 5 2 4 1 Battery Installation Replacement 6 2 4 2 Power Up 6 2 4 3 Detector Connection and Setup 6 2 4 4 Setting the Detector Calibration Factor 7 2 4 4 1 Setting the Calibration Facto...

Page 6: ...2 Performance Verification 14 4 2 1 Environmental Conditions 14 4 2 2 Recommended Test Equipment 14 4 2 3 Initial Conditions 15 4 2 4 Verification Procedures 15 4 3 Troubleshooting Guide 17 Section 5...

Page 7: ...iagram 10 Figure 7 Model 1815 C Range Switch Positions 15 List of Tables Table 1 Model 1815 C Compatible Detectors 2 Table 2 Model 1815 C Controls 3 Table 3 SETUP DIP Switch Functions 5 Table 4 SETUP...

Page 8: ...able to this instrument Warning The American National Safety Institute ANSI states that a shock hazard exists when probes or sensors are exposed to voltage levels greaterthen42VDCor42VpeakAC Donotexce...

Page 9: ...pe C degrees Centigrade DC direct current F degrees Fahrenheit Hz hertz cycles per second I V current to voltage converter kHz kilohertz k kiloOhm mA milliamp mV millivolt nA nanoamp nF nanofarad nm n...

Page 10: ...5 Digit LCD Sample Rate 2 5 Hz Range Switch 6 positions one per decade of gain Operating Environment 70 RH Noncondensing 18 C 28 C 65 F 82 F Storage Environment 90 RH Noncondensing Compatible Detector...

Page 11: ...the calibration factor in the following way Display Noise Floor specified noise floor calibration factor mantissa Voltage Amplification High Power Detectors Range Switch Position 3 4 5 6 High Power Fu...

Page 12: ...floor is greater than specified see Section 3 3 2 Quantum Detector Tempera ture Effects 6 ACCEL ADJ fully clockwise worst case Detector Responsivity Limits Detector in Use Allowed Responsivities Allo...

Page 13: ...xii...

Page 14: ...FP3 FH1 818T 10 818T 30 818T 150 818T 150X 818T 300 High Power Low Power W mW W W W W W mW mW mW W W W kW 20 200 2 2 200 20 nW nW LOW POWER HIGH POWER LOW POWER W ATTN CAL ADJ CAL POWER ZERO ON 1 OFF...

Page 15: ...del 1815 C Optical Power Meter Refer to Figure 1 for where each accessoryfitswithintheModel1815 COpticalPowerMeterfamilytree For moreinformation pleaserefertotheNewportCatalog Table1 Model1815 CCompat...

Page 16: ...The front and rear panels of the Model 1815 C are shown in Figures 2 and 3 respectively ThefollowingtablegivesashortdescriptionofeachModel 1815 Ccontrol Table2 Model1815 CControls Control Description...

Page 17: ...Adj Reqd See Manual Cal Factor Exp Low Power Low Power w Attn High Power INPUT 1 0 ANALOG OUTPUT ACCEL ADJ DIP Switch Input BNC Connector Acceleration Time Constant Adjust Output BNC Connector SETUP...

Page 18: ...h Thefollowingtableidentifiesthefunc tioneachSETUPswitchpositionperforms Table3 SETUPDIPSwitchFunctions SETUPSwitch SwitchPosition Description 1 2 SelectsforLow Power semiconduc tor detector 1 0 orHig...

Page 19: ...themeter Thedigitaldisplayshouldlightindicatingthatthemeteris operating IftheBATannounciatorislit replacethebatteries SeeSection 2 4 1 2 4 3 Detector Connection and Setup Connect the detector to the I...

Page 20: ...ed ddofthefullcalibrationfactor valued ddE d Themantissaisthemostfrequentcalibrationadjustment madewhenmakingmeasurementsatvariouswavelengths Tosettheman tissa usethefollowingprocedure 1 Pressthefront...

Page 21: ...econdtimeconstant theACCELADJ potentiometerisadjustedatthefactoryforthisvalue However becausethe actualtimeconstantisnotexactly1secondforallthesmallerthermopiles it mayberequiredtooptimizetheaccelerat...

Page 22: ...no adjustment to the ACCELADJisrequired Note AccelerationcanonlybeusedwithHigh Power thermopile detec tors Ifaccelerationisusedwithotherdetectors erroneousreadings canresult 2 4 6 Analog Output The Mo...

Page 23: ...ordiskthermopiles voltagemeasurement DetectorsignalsareintroducedtotheModel1815 CbywayofaBNCinput connector A bank of DIP switches found on the back of the Model 1815 C configurestheamplifieraseithera...

Page 24: ...ion Low Powerdetectors aresignificantlyaffectedbytemperature Atlonger wavelengths thesequantumdetectorstypicallylosesensitivitywithincreasing temperature Howeverthedetectordarkcurrentincreasesexponent...

Page 25: ...Powerdiskthermopiledetectors whichisnotcomingfromthesource andturningoffroomandotherlights Note Changes in ambient light levels can occur from such factors as turning room lights on or off or by movin...

Page 26: ...easurementwas made WavelengthCalibration Any CheckthattheCalibration Factorforthemeasure mentwavelengthis properlyset LowBatteryPower Any Checkthatmeterisnot indicatinglowBAT DetectorSaturation Low Po...

Page 27: ...ion Note If the instrument does not meet specifications and it is still under warranty lessthan12monthssincedateofshipment contactyour local Newport Corporation representative or the factory to deter...

Page 28: ...gapreciselysetablecurrentsourcesuchasthe KeithleyInstrumentsModel220andapreciselysetableVoltageSourcesuchas theHewlett Packard8904A FunctionGenerator W W kW W W mW W nW nW mW W W W 2 200 20 20 200 2 m...

Page 29: ...htoposition3 thehighestgainpositioninthe voltage mode See Figure 7 3 PresstheCALbuttonandadjustCALADJuntilthedisplayindicates 1 00andthenreleasetheCALbutton 4 With a shorted input use the ZERO knob to...

Page 30: ...hissection pleaseseeSection5fordetails onreturningyourentiresystemtoNewportforservice Table10 Symptom FaultTroubleshootingGuide Symptom PossibleFault Correction Blankdisplay PowerswitchOFF TurnswitchO...

Page 31: ...anbatteries see Section2 4 1 Itscalibrationaccuracyiswarrantedforaperiodof1year After 1year theunitshouldbereturnedtoNewportCorporationforrecalibration Toobtaininformationconcerningfactoryservice cont...

Page 32: ..._________________________________________ Company ______________________________________________________________________________ Address _______________________________________________________________...

Page 33: ...20...

Page 34: ...en certified compliant with ISO 9001 by the British Standards Institution P N 19571 01 Rev F IN 03923 1 00 Printed in the USA Newport Corporation Worldwide Headquarters 1791 Deere Avenue Irvine CA 926...

Reviews: