Section 2
System Operation
This section contains the information needed to prepare and operate the
Model 1815-C Optical Power Meter. Operation consists of using the Model
1815-C to perform basic optical power measurements using one of Newport’s
compatible detectors.
Operating controls and settings for the Model 1815-C are found on both the
front and rear panels. For general operation, only the front panel controls are
used. The rear panel controls and settings are used for meter/detector setup.
The front and rear panels of the Model 1815-C are shown in Figures 2 and 3
respectively. The following table gives a short description of each Model
1815-C control:
Table 2 — Model 1815-C Controls
Control
Description
Comment
POWER
Push Button
Turns the Model 1815-C on and off
CAL
Push Button
Causes the meters CAL value to be
displayed.
CAL ADJ
10 Turn Trim Pot
Sets the mantissa of the Calibration
Factor.
ZERO
10 Turn Knob
Zeros the display. Used to elimi-
nate ambient signals.
Range
6 Position Knob
Adjusts the meter’s signal gain.
ACCEL ADJ
25 Turn Trim Pot
Sets the thermopile acceleration
time constant.
SETUP
4 Switch DIP Bank
Configures the meter to the
detector in use.
Introduction
2.1
Front and Rear Panels
2.2