30 • Computer Control
NEW FOCUS, Inc.
Command Definitions
System Status Commands
Identification Query
Syntax
*
IDN?
Description
Returns the system identification string containing
the manufacturer, model number, serial number,
and firmware revision numbers for the laser cavity
and controller.
Example
*IDN?
⇒
New Focus Inc., 6427, 02149, H1.67, C1.00
(Manufacturer = New Focus, Model= 6427, Serial
Number=02149, Motor Firmware Rev. 1.67, Con-
troller Firmware Rev. 1.00)
Operation Complete Query
Syntax
*OPC?
Description
The laser has four long-term operations:
•
Turning on the system (system initialization)
•
Turning on the diode (:OUTPut ON)
•
Scanning (:OUTPut:SCAN:START)
•
Set Wavelength ([:SOURce]:WAVElength)
Starting any of these operations clears the OPC sta-
tus bit. When the laser returns to the start wave-
length or reaches the target wavelength, the OPC
status bit is set to 1. The bit is also set to 1 if the
movement is interrupted, either with
OUTPutSCAN:ABORt, which interrupts a scan, or
OUTPut:TRACk OFF, which interrupts a wave-
length set.
Argument/Response
Returns 0 if a long-term operation is ongoing.
Returns 1 if no long-term operation is ongoing.
64X7 rev E.fm Page 30 Tuesday, April 3, 2001 4:16 PM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com