10 • Getting Started
NEW FOCUS, Inc.
The laser is shipped with brackets for mounting the system in a rack
and with rubber feet for using the system on a table top. Use a 1/8"
Allen wrench to add or remove the feet or the brackets.
Starting the Laser for the First Time
The following section takes you through the basic steps of starting up
and shutting down the laser. The controls and functions are described
in more detail in the following chapter.
Figure 3:
Front Panel
1. Make sure the laser aperture is blocked or attached to an opti-
cal fiber that is connected to an appropriate receptacle.
Only
use fiber with FC/APC connectors.
2.
Check the voltage setting:
Check that the power module on
the back of the laser is set for the proper AC line voltage (see
page 51 for information on checking and changing the voltage).
3. Connect the power cord:
Attach the power cord to the laser and
plug it into a wall outlet.
4. Turn on the system:
Turn on the
AC Power
keyswitch (position
“
1
”) to initialize the system: the system ID will appear in the
wavelength display during initialization.
After turning the keyswitch, allow the system a minimum of 45 minutes to
warm up before turning on the laser diode (step 6.) to ensure the system meets
its operating specifications.
Once you turn on the keyswitch and the system initializes (about
30 seconds), you can operate the system remotely through the
Addressed
Remote
Laser Power
AC Power
0
1
Local
Wavelength
Adjust
Track
Wavelength (nm)
Current (mA)
Laser Output
FC/APC
Power
Adjust
Power Setpoint (dBm)
64X7
Made in USA
AVOID EXPOSURE
INVISIBLE LASER RADIATION
EMITTED FROM
THIS APERTURE
INVISIBLE LASER RADIATION AVOID
DIRECT EXPOSURE TO BEAM
Power output
Wavelength
CLASS IIIB LASER PRODUCT
Start
λ
Stop
RS 232
GPIB
Step
Hi Res
Tunable Laser Source
V
IDIA™
-D
ISCRETE
Note:
Note:
64X7 rev E.fm Page 10 Tuesday, April 3, 2001 4:16 PM