Trouble
(internal failure)
On display
Green
POWER
LED
Amber
TROUBLE
LED
Red
ALARM
LED
Remarks
(1)
Low flow rate
FLOW
On
Flashing
Off
Flow-rate icon stops
spinning.
(2)
Sensor fault
SENS.
On
Flashing
Off
(3)
Incorrect sensor
installed
SENS.
On
Flashing
Off
―- - - -‖ shown for the
gas concentration on
the LCD.
(4)
Broken pyrolyzer
wire
CONV.
On
Flashing
Off
(5)
No power
Off
Off
Off
(6)
Blown fuse
Off
Off
Off
(7)
Memory read
error
nG
On
Flashing
Off
Error code shown at
the bottom-right corner.
CAUTION
Oxygen sensor (COS-7): The analog output from this product falls to 0.6 mA or less in the event
of a trouble alarm as listed above. When the host system (e.g., control room) is set to the low
limit alarm (gas alarm activates when the gas concentration falls to the alarm set value or less)
and if the product
’s analog output falls to 0.6 mA or less in under a second without a gas alarm, it
means that a trouble alarm is present. Take measure to prevent the host system from activating
a gas alarm, as needed.