Poly-Spectrum-8/V (Technical Manual) 

 

20 

3. Proper Use 

3.1.  Getting Ready 

Operating Limitations: 

  Ambient temperature is from +10 to +35

С. 

  Relative humidity is from 30 to 85% (non-condensing). 

  Atmospheric pressure is from 70 to 106 kPa. 

Before  the  power  supply  is  switched  on,  make  sure  that 
the electronic  unit  and  computer  equipment  cases  have  no 
apparent mechanical failures which can represent danger. 

Power Supply Switch on and System Test 

The power supply of the system is switched on by pressing the 

Power

 button on the PC. The electronic unit has no power sup-

ply and it is connected to the PC all the time. The power supply 
is  switched  on  when  the  operating  system  is  loaded  and  the 

Poly-Spectrum.NET

 program is started. 

3.2.  Getting Started 

Before  starting  ECG  exams,  configure  the  system  taking  into 
consideration  exam  type  and  recommendations  of  the  user 
manual for this software. 

The examination includes the following stages: 

  preparation  of  animal  for  ECG  recording  and  placement  of 

electrodes; 

  recording of ECG signals; 

  analysis of the obtained results and their printing. 

Summary of Contents for Poly-Spectrum-8/V

Page 1: ...Technical Manual Poly Spectrum 8 V Veterinary Digital ECG System TM003 04 001 000 01 07 2019...

Page 2: ...Neurosoft 2020 5 Voronin str Ivanovo 153032 Russia P O Box 10 Ivanovo 153000 Russia Phone 7 4932 95 99 99 7 4932 24 04 34 Fax 7 4932 24 04 35 E mail info neurosoft com Internet www neurosoft com...

Page 3: ...tallation 16 2 1 Requirements to Personnel 16 2 2 Room Selection and Placement 16 2 3 Unpacking and Check of Delivery Set 17 2 4 System Assembly and Connection 18 3 Proper Use 20 3 1 Getting Ready 20...

Page 4: ...4 6 Disposal 30 7 Delivery Set and Package Data 30 8 Warranty 30 9 Reclamation 32 Annex 1 Delivery Set 33 Annex 2 Electromagnetic Emissions and Immunity 35...

Page 5: ...the following address Post Office Box 10 Ivanovo 153000 Russia or by e mail help neurosoft com You can find additional information about Neurosoft products on our website www neurosoft com or ask que...

Page 6: ...esearch field General Description The system is intended for veterinary use only The Poly Spectrum 8 V system is used to estimate the electrical activity of the heart muscle and cardi ac performance d...

Page 7: ...may occur Safety Measures To provide safety measures and exclude the possibility of electric trauma of medical staff or examined animal it is PROHIBITED to use the system which mounting and setting w...

Page 8: ...g into liquid autoclaving or steam cleaning as this may damage equipment or reduce its usable life The disinfection of the system is described in sec tion 4 3 Disinfection drop the system from more th...

Page 9: ...60 up to 75 Hz from 10 up to 5 from 30 up to 5 Relative error of interval measurement within 7 Relative error of calibration pulses recording within 5 Heart rate measurement range 30 240 beats per min...

Page 10: ...e NS017103 008 2 5 0 2 2 0 0 1 1 5 0 1 1 45 0 1 Weight electronic unit transportation bag packed system without computer 0 3 0 05 kg 0 15 0 05 kg not more than 1 5 kg Applied parts CF type with protec...

Page 11: ...of this technical manual may result in increased emission and system de creased immunity As for safety the system satisfies IEC 60601 1 2012 AAMI ANSI ES 60601 1 2005 R2012 and A1 2012 and A2 2010 R...

Page 12: ...channel analog to digital converter ADC The central processor refers to the ADC from time to time pro cesses the received information and transfers it to the USB processor The USB processor transfers...

Page 13: ...hown in Fig 2 and Fig 3 Fig 2 Front panel of Poly Spectrum 8 V If the LED indicator is orange the device is connected to the computer and 5 V voltage is supplied to USB The green color of the LED indi...

Page 14: ...gainst defibrillator discharge according to IEC 60601 1 2012 AAMI ANSI ES 60601 1 2005 R2012 and A1 2012 and A2 2010 R 2012 EN 60601 1 2006 A1 2013 and IEC 60601 1 2 2014 EN 60601 1 2 2015 and IEC 606...

Page 15: ...of the Member State of the Eurasian Economic Community and the Customs Union ingress protection according to IEC EN 60529 The equipment is identified with the GS1 128 barcode integrat ed to the barcod...

Page 16: ...consideration the power wiring and protective ground in the room Please read the following requirements and recommen dations Requirements concerning the room selection and equip ment placement The loc...

Page 17: ...cked by electrician The recommended distance from the electronic unit to the nearest electric mains is not less than 3 meters 2 3 Unpacking and Check of Delivery Set In case the box with the system wa...

Page 18: ...program is described in the user manual If the distributive is missing or the software update is re quired address to your local dealer The authorized Neurosoft dealers are listed on the website https...

Page 19: ...Assembly and Installation 19 Fig 6 Configuration 1 Fig 7 Configuration 2...

Page 20: ...power supply of the system is switched on by pressing the Power button on the PC The electronic unit has no power sup ply and it is connected to the PC all the time The power supply is switched on wh...

Page 21: ...o the body It is desirable that the limbs are located parallel The electrodes are applied in the area of metacarpus and metatarsus directly on the skin pre treated with alcohol and moistened with cond...

Page 22: ...ical Manual 22 The placement of electrodes for X Y lead is shown in Fig 9 Fig 9 Placement of electrodes for X Y lead The placement of electrodes for apex lead is shown in Fig 10 Fig 10 Placement of el...

Page 23: ...ed with alcohol and moistened with conductive fluid electrode gel It is not necessary to cut out the wool in the areas of electrode place ment except for long term cardiac monitoring In practice it is...

Page 24: ...in the obtained record The analyzed part of the record should not be less than 1 mi nute If arrhythmia or ectopic beats are detected the recording should be increased to 20 30 minutes 3 3 Troubleshoo...

Page 25: ...e appears on the screen 1 The database directory is not available 2 The database directory has been renamed or replaced 1 Check the PC connection to the local network Make sure that the PC with the da...

Page 26: ...the information given in sec tion 3 3 Troubleshooting If the troubles can t be eliminated using the control tools of the system or by restart it should be switched off and checked by the specialist Th...

Page 27: ...ng agent scrubbing pads or other abrasive appli cators Keep all cleaning fluids away from electrical connectors Visually inspect the interface cables and power cords that are used with components and...

Page 28: ...urrent repair of the system includes the repair of some component parts and cables If the component parts are con nected to the system the repair is prohibited When performing the repair all units mus...

Page 29: ...examined externally and the circuit is checked for short circuit or break In case the cable is broken replace it or cut it if the cable length is suffi cient Please pay attention for the cable labeli...

Page 30: ...ion about the delivery set is described in the package report which is an integral part of this document and should be kept along with it 8 Warranty 8 1 The manufacturer guarantees the system quality...

Page 31: ...f charge The repair is carried out in Neurosoft service center 5 Voronin str Ivanovo 153032 Russia in compliance with the procedure stated in chapter 9 Reclamation You can also contact the Authorized...

Page 32: ...res If possible please indicate the reasons and circumstances preceding the fault detec tion in addition it is recommended to add the test report the test data photos and other materials allowing solv...

Page 33: ...Technology China NS003103 009 1 Veterinary ECG electrode clip VV02 Shenzhen Medke Technology Co Ltd China 8 Reusable plate electrode 20x15 2 F9024 F9024 RU10P SSC FIAB Italy 3 4 2 Reusable plate elect...

Page 34: ...with Poly Spectrum NET Veterinary additional module 1 Operational Documentation Poly Spectrum 8 V technical manual TM003 04 001 000 1 Poly Spectrum NET user manual with Poly Spectrum NET Veterinary a...

Page 35: ...environment guidance RF emissions CISPR 11 Group 1 The system uses RF energy only for its internal function Therefore its RF emissions are very low and not likely to cause any interference in nearby e...

Page 36: ...5 kV 1 kV2 Mains power quality should be that of typical commercial or hospital environment 2 kV common mode 0 5 kV 1 kV 2 kV2 Voltage dips IEC 61000 4 11 2004 0 UT during 0 5 period at 0 45 90 135 1...

Page 37: ...ecommended separation distance calculated from the equation applicable to the frequency of the transmitter The recommended separation distance Conducted RF IEC 61000 4 6 2013 3 V in 0 15 MHz 80 MHz ba...

Page 38: ...eld strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted t...

Page 39: ...ance according to frequency of transmitter d m P d 17 1 150 kHz up to 80 MHz outside ISM bands P d 17 1 80 MHz to 800 MHz P d 23 2 800 MHz to 2 5 GHz 0 01 0 117 0 117 0 233 0 1 0 369 0 369 0 738 1 1 1...