Description
11
Safety and Electromagnetic Compatibility
Electromagnetic compatibility (EMC) is provided by conform-
ance to IEC 60601-1-2-2014 (EN 60601-1-2:2015) require-
ments.
The system is intended for operation in electromagnetic envi-
ronment, which special features are specified in Annex 2.
Portable and mobile RF communication equipment can affect
the system operation.
The use of equipment not listed in Table 3 of this technical
manual may result in increased emission and system de-
creased immunity.
As for safety, the system satisfies IEC 60601-1:2012
(AAMI/ANSI ES 60601-1:2005/(R2012) and A1:2012 and
A2:2010/(R)2012, EN 60601-1:2006/A1:2013) and IEC 60601-
1-2-2014 (EN 60601-1-2:2015) and IEC 60601-2-25:2011 (EN
60601-2-25:2015).
The electronic unit is supplied by regulated power supply
through USB interface, it has double isolation and CF type ap-
plied parts with protection against defibrillator discharge accord-
ing
to
IEC
60601-1:2012
(AAMI/ANSI
ES
60601-
1:2005/(R2012) and A1:2012 and A2:2010/(R)2012, EN 60601-
1:2006/A1:2013).