Maintenance
29
4.
Maintenance
4.1. General Requirements
The safety measures when servicing conforms to the ones described in chapter 3
“Proper Use”.
The qualification requirements to the staff are listed in section 2.1
“Requirements to
Personnel”.
The maintenance of the bought articles included in the system is conducted according
to user and technical manuals or typical rules.
When detecting the troubles, use the information given
in section 3.3 “Troubleshoot-
ing”. If the troubles can’t be eliminated using the control tools of the system or by re-
start, it should be switched off and checked by the specialist.
The type, amount and intervals of maintenance, except the ones specified in this sec-
tion, are not specifically established.
The check of delivery set is done by conformity to the packing report for the system.
4.2. User Maintenance
The system maintenance in the process of operation includes the external examina-
tion, check of connectors and cables, removal of contaminations from the surface of
the cases using wet fabric, and also disinfection according to section 4.3.
4.3. Disinfection
Before cleaning the electronic unit, switch it off. As you clean, visually inspect the unit
and its components for damage or wear. Contact
Neurosoft
if you notice damage to the
exterior of the component.
For routine cleaning of the electronic unit, use a cloth gently wrung in phenoles (Bacil-
lotex® etc.) or 70% alcohol, 0,5% chlorohexidine.
If dangerous virus contamination is suspected,
use aldehydes (Cidex® etc.) or chlorin-
ates (Di
versol BX®).
Be careful not to drip disinfectant directly into the input and output plugs and other open-
ings in the cover. Remove disinfectant with a dry cloth. Do not use abrasive or solvent
silicon-based cleaning agent, scrubbing pads or other abrasive applicators.
Keep all cleaning fluids away from electrical connectors.
Visually inspect the interface cables and power cords that are used with components
and accessories. If you notice unusual wear or breakage, disconnect the cable or cord
immediately, and contact
Neurosoft
for replacement. Gently wipe them with a soft
cloth moistened with disinfectant (for example, 1% chloramine solution or 3% hydro-