Neuro-ERG/V (Technical Manual)
16
Fig. 8. The acquisition start window from external stimulator.
4.
In the stimulation settings it is recommended to set those stimulus parameters
which the external stimulator has as these particular values are saved together
with the trace and taken into analysis.
5.
Run the
Acquisition|Acquisition/stimulus
menu command. The device goes to the
external stimulus standby mode. If this menu command is not executed, any
stimuli from the external stimulator are ignored.
6.
Start stimulation from the external stimulator.
In other aspects the signal acquisition does not differ from the one described in
the corresponding chapters of the user manual.
The external stimulators connected to the trigger output must be used as follows:
1.
Connect the external stimulator to the trigger output using a cable.
2.
Switch on the external stimulator power supply.
3.
Run
Neuro-MEP.NET
software.
4.
In the stimulation settings it is recommended to set those stimulus parameters
which the external stimulator has as these particular values are saved together
with the trace and taken into analysis.