2.5.2 Alarm Mode
The UniMano has an audio beeper that always activate during a overpressure
situation, for low battery situation and after Leak Detection. If you press [F2]
in the Setup Menu and select ‘ON’ [F1] the UniMano will also beep when
zeroing the unit and if Peak Function is selected. Pressing ‘OFF’ [F2] will
deactivate these two alarms. The UniMano will default on initiation to the
‘OFF’ position.
Beeper OFF
ON
OFF
2.6
SIMPLE TROUBLESHOOTING GUIDE
When inaccurate pressure vacuum readings are suspected, check the following items:
Install a fresh 9 Volt battery.
Insure the display reads zero with no pressure applied to the instrument.
Auto zero if necessary.
Apply a known pressure, and verify the reading within the specifications of the
UniMano 1000 model, 0.1% Full Scale
2.7
TYPICAL APPLICATION SET UP
UniMano 1000 Series
13