background image

NEOGEN.com

12

CHAPTER FIVE

User Access

Logging In

The user login information is set up and administered through the data manager software.For information on setting 
up users, see the user management section of the AccuPoint

®

 data manager user guide. By default, when no users are 

loaded to the device, the login screen will be bypassed, and the user will automatically be logged in as a guest user. 
When users have been loaded in the device, users will be prompted to log in to the device before running tests. The 
username of the logged-in user will be recorded on any tests taken in that session. 
To log in, select the username field. Navigate to the appropriate username in the list by scrolling with the up and down 
buttons on the d-pad or use the search field on the display to search for a specific username. Select the appropriate 
username and proceed to PIN entry. If a PIN was established for the user, enter the PIN in the PIN field, and select sign 
in to continue. If no PIN was set for the user, select sign in to continue.

Logging Out

The username of the logged-in user will be displayed at the top right of the screen. To log out, select the user icon and 
select sign out. Instrument will return to the log in screen and a user will need to log in before new samplers can be read 
or other instrument features can be accessed.

Summary of Contents for AccuPoint Advanced Next Generation

Page 1: ...AccuPoint Advanced Next Generation User Guide...

Page 2: ...of the Instrument 06 Cleaning the Instrument Chapter 4 Preparing the Instrument for Use 07 Power 08 Home Page 09 Running Quick Tests 09 Results Screen 09 Add Notes 09 Retest 10 Sample Procedure Chapt...

Page 3: ...ager Software 20 Sampler Type Surface Liquid or Alkaline Phosphatase 20 Setting the Pass Marginal and Fail Thresholds 21 Troubleshooting Chapter 11 Technical Information 25 Electromagnetic Compatibili...

Page 4: ...t of providing greater sensitivity through a clearer distinction between very low levels of ATP The AccuPoint Advanced NG instrument offers enhanced navigation touch screen wireless data transfer capa...

Page 5: ...a food debris yeast and mold Bioluminescence is a chemical reaction that produces light ATP bioluminescence occurs when ATP from a sample comes into contact with luciferase an enzyme found in fireflie...

Page 6: ...NEOGEN com 4 Eject button Full color touch screen display USB data and charging port RFID antenna Sampler compartment On Off and navigation d pad Operating Principles and Technology...

Page 7: ...esting liquids Be sure to remove and properly dispose of samplers after each reading from the instrument The sampler reagents and their cartridges are non toxic and may be disposed of without any spec...

Page 8: ...at may be contaminated To clean the lens use the AccuPoint Advanced NG cleaning swab moistened with a solution of isopropyl alcohol 70 Be sure to turn the instrument upside down while cleaning the len...

Page 9: ...too low a notification will appear recommending you charge the instrument immediately Once the battery level reaches a critically low state a notification will appear warning that the instrument is p...

Page 10: ...e AccuPoint Advanced NG instrument connects to the data manager software on your PC Battery Icon indicates the amount of battery life left on the instrument Menu Navigation menu with sites results set...

Page 11: ...the last used site and will not initiate a quick test Selecting quick test will always use the quick test site A progress indicator will appear displaying the progress of the sample Once complete the...

Page 12: ...ampling Under standard testing select the correct site to be tested 03 Sample your site by drawing a 4 x 4 in 10cm x 10cm square With a vertical and horizontal back and forth motion fill in the square...

Page 13: ...e sampler and begin test Note keep instrument upright while the sample is being read 08 The RLU value will display along with a symbol for pass marginal or fail 09 On the results screen users have the...

Page 14: ...ny tests taken in that session To log in select the username field Navigate to the appropriate username in the list by scrolling with the up and down buttons on the d pad or use the search field on th...

Page 15: ...thout the use of a plan subset proceed to area selection To run a standard test as a part of a plan subset select select plan Navigate to the appropriate plan in the plans list by scrolling with the u...

Page 16: ...ure can be used to avoid confusion and save time by automatically pulling up the appropriate group based on the location of a physical RFID tag See the RFID section for more information RFID tags are...

Page 17: ...tton to confirm the added note Notes will be stored in the comments field when results are imported into the data manager software Retest To run a retest of the same site remove sample and select the...

Page 18: ...ll dim when the instrument is idle Off Timeout Adjusts the amount of time before the instrument automatically powers off when the instrument is idle Auto Select Site Automatically selects the next sit...

Page 19: ...data transfer to occur A Wi Fi icon with a lock will appear when the instrument is connected to Wi Fi and ready to transfer To sync results select sync results to transfer test results wirelessly to...

Page 20: ...rate when it recognizes the RFID tag A green check mark will appear next to groups which have RFID tags assigned Be sure to note on the RFID sign which group has been associated with it This will help...

Page 21: ...munication the sync results button can be used to send new readings data to the active data manager database When the sync results button is pressed the device will attempt to re establish a secure co...

Page 22: ...two sampling modes surface and liquid Both the AccuPoint surface and access samplers are designed to be read in surface mode The AccuPoint liquid sampler is read in liquid mode A water droplet symbol...

Page 23: ...NEOGEN com 21 Troubleshooting Error 01 Door open during unit initialization while powering on Error 02 Try to run test when unit is being charged Additional Features...

Page 24: ...NEOGEN com 22 Error 03 Reminder to remove used sampler after testing is complete Error 04 Please wait baseline in progress Additional Features...

Page 25: ...NEOGEN com 23 Error 05 Leaving the door open when the unit is trying to take a baseline reading Error 06 The unit battery is low charge unit soon by connecting to the USB charger Additional Features...

Page 26: ...07 The unit battery is depleted the unit is shutting down Connect the unit to the USB charger Error 08 Tilt warning flag unit was not kept upright during the test Test results are marked with this fla...

Page 27: ...andard s Operationissubjecttothefollowingtwoconditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation...

Page 28: ...this product There is no warranty of merchantability of this product or the fitness of the product for any purpose NEOGEN shall not be liable for any damages including special or consequential damage...

Page 29: ...NEOGEN Corporation 2021 NEOGEN and AccuPoint are registered trademarks of NEOGEN corporation Lansing MI 48912 U S NEOGEN com...

Reviews: