Performance Considerations
O
XI
Max
N-65 Operator’s Manual
91
To prevent interference from ambient light, ensure that
the
O
XI
M
AX
sensor is properly applied, and cover the
O
XI
M
AX
sensor site with opaque material.
If interference due to patient activity presents a problem,
try one or more of the following to correct the problem:
•
verify that the
O
XI
M
AX
sensor is properly and
securely applied
•
move the
O
XI
M
AX
sensor to another site
•
use an adhesive
O
XI
M
AX
sensor
•
use a new
O
XI
M
AX
sensor with fresh adhesive
backing
•
keep the patient still, if possible
If interference due to poor perfusion presents a problem,
consider using the MAX-R
O
XI
M
AX
sensor or the MAX-
FAST
O
XI
M
AX
sensor. The MAX-R
O
XI
M
AX
sensor
obtains measurements from the nasal septal anterior
ethmoid artery, an artery supplied by the internal carotid.
These
O
XI
M
AX
sensors may obtain measurements when
peripheral perfusion is relatively poor.
Summary of Contents for OxiMax N-65
Page 1: ......
Page 12: ...Safety Information 6 OXIMax N 65 Operator s Manual ...
Page 14: ...Introduction 8 OXIMax N 65 Operator s Manual ...
Page 36: ...Set up the OXIMAX N 65 30 OXIMax N 65 Operator s Manual ...
Page 92: ...OXIMAX Sensors and Accessories 86 OXIMax N 65 Operator s Manual ...
Page 98: ...Performance Considerations 92 OXIMax N 65 Operator s Manual ...
Page 108: ...Troubleshooting 102 OXIMax N 65 Operator s Manual ...
Page 112: ...Maintenance 106 OXIMAX N 65 Operator s Manual ...
Page 118: ...Menu Structure 112 OXIMAX N 65 Operator s Manual ...
Page 150: ......
Page 151: ......