![Nellcor OxiMax N-65 Operator'S Manual Download Page 35](http://html1.mh-extra.com/html/nellcor/oximax-n-65/oximax-n-65_operators-manual_718076035.webp)
Set up the OXIMAX N-65
O
XI
Max
N-65 Operator’s Manual
29
•
failure to cover the
O
XI
M
AX
sensor site with opaque
material in high ambient light conditions
•
venous pulsation
•
dysfunctional hemoglobin
•
low perfusion
Loss-of-pulse signal can occur for the following reasons:
•
the
O
XI
M
AX
sensor is applied too tightly
•
defibrillation
•
a blood pressure cuff is inflated on the same
extremity as the one with the
O
XI
M
AX
sensor
attached
•
there is arterial occlusion proximal to the
O
XI
M
AX
sensor
•
poor peripheral perfusion
•
loss of pulse/cardiac arrest
High ambient light sources that can interfere with the
performance of an SpO
2
O
XI
M
AX
sensor are:
•
surgical lights (especially those with a xenon light
source)
•
bilirubin lamps
•
fluorescent lights
•
infrared heating lamps
•
direct sunlight
Summary of Contents for OxiMax N-65
Page 1: ......
Page 12: ...Safety Information 6 OXIMax N 65 Operator s Manual ...
Page 14: ...Introduction 8 OXIMax N 65 Operator s Manual ...
Page 36: ...Set up the OXIMAX N 65 30 OXIMax N 65 Operator s Manual ...
Page 92: ...OXIMAX Sensors and Accessories 86 OXIMax N 65 Operator s Manual ...
Page 98: ...Performance Considerations 92 OXIMax N 65 Operator s Manual ...
Page 108: ...Troubleshooting 102 OXIMax N 65 Operator s Manual ...
Page 112: ...Maintenance 106 OXIMAX N 65 Operator s Manual ...
Page 118: ...Menu Structure 112 OXIMAX N 65 Operator s Manual ...
Page 150: ......
Page 151: ......