Specifications
134
O
XI
Max
N-65 Operator’s Manual
Table 11: Electromagnetic Immunity
The
O
XI
M
AX
N-65 is intended for use in the electromagnetic
environment specified below. The customer or user of the monitor
should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
+ 6 kV contact
+ 8 kV air
Complies
Floors should be
wood, concrete, or
ceramic tile. If
floors are covered
with synthetic
material, the
relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC 61000-4-4
+ 2 kV power
supply lines
+ 1 kV for
input/output
lines
Complies
Main power should
be that of a typical
commercial or
hospital
environment.
Power
Frequency
(50/60 Hz)
magnetic field
3 A/m
Complies
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Summary of Contents for OxiMax N-65
Page 1: ......
Page 12: ...Safety Information 6 OXIMax N 65 Operator s Manual ...
Page 14: ...Introduction 8 OXIMax N 65 Operator s Manual ...
Page 36: ...Set up the OXIMAX N 65 30 OXIMax N 65 Operator s Manual ...
Page 92: ...OXIMAX Sensors and Accessories 86 OXIMax N 65 Operator s Manual ...
Page 98: ...Performance Considerations 92 OXIMax N 65 Operator s Manual ...
Page 108: ...Troubleshooting 102 OXIMax N 65 Operator s Manual ...
Page 112: ...Maintenance 106 OXIMAX N 65 Operator s Manual ...
Page 118: ...Menu Structure 112 OXIMAX N 65 Operator s Manual ...
Page 150: ......
Page 151: ......